Detector Background And Sensitivity Of Semiconductor X-Ray Fluorescence Spectrometers

Published online by Cambridge University Press:  06 March 2019

F. S. Goulding
Affiliation:
Lawrence Berkeley Laboratory University of California Berkeley, California 94720
J. M. Jaklevic
Affiliation:
Lawrence Berkeley Laboratory University of California Berkeley, California 94720
B. V. Jarrett
Affiliation:
Lawrence Berkeley Laboratory University of California Berkeley, California 94720
D. A. Landis
Affiliation:
Lawrence Berkeley Laboratory University of California Berkeley, California 94720
Get access

Abstract

Degraded detector pulses are shown to contribute most of the background in spectra produced by semiconductor-detector X-ray spectrometers. A new guard-ring detector is used with appropriate circuits to reduce background by a large factor. We discuss the sensitivity of the new arrangement with various excitation sources.