Window Area Effects in the Detector Efficiency for Source Excited EDXRF Geometries

Published online by Cambridge University Press:  06 March 2019

J.M.F. dos Santos
Affiliation:
Physics Department University of Coimbra 3000 Coimbra Portugal
Benilde E.O. Costa
Affiliation:
Physics Department University of Coimbra 3000 Coimbra Portugal
Get access

Abstract

The efficiency dependence on detector window area for three different geometries of source excited energy dispersive X-ray fluorescence (EDXRF) analysis systems are calculated using the Monte Carlo method. Two typical samples (coal and oil) are considered for discussion. The results show that the efficiency is affected mainly by geometric factors: source-to-sample distance and absorption of the exciting X-rays; fluorescent photon self-absorption has a much smaller effect.