The measurement of blood pressure

Published online by Cambridge University Press:  17 February 2009

J. C. Patterson
Affiliation:
Department of Mathematics, University of Queensland, St. Lucia, QLD. Australia.
Rights & Permissions [Opens in a new window]

Abstract

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the 'Save PDF' action button.

In this paper, the procedure of the clinical measurement of blood pressure is modelled by the application of a uniform pressure band to a long, homogeneous, isotropic cylinder. The deformations are assumed to be infinitesimal, and transform methods are used to analyse the resulting equations. The inversion of the resulting transforms is carried out numerically. It is shown that, in spite of the fairly crude assumptions of the model, the actual load on the artery may be markedly different from that applied to the surface, leading to inaccuracies in the measured blood pressure. The parameter of importance is shown to be the ratio of pressure band width to arm diameter.