Bacteriological evaluation of a mobile laminar cross-flow unit for surgery, under laboratory circumstances

Published online by Cambridge University Press:  15 May 2009

D. van der Waaij
Affiliation:
Radiobiological Institute TNO, Rijswijk, The Netherlands
N. Wiegersma
Affiliation:
Radiobiological Institute TNO, Rijswijk, The Netherlands
J. Dankert
Affiliation:
Department for Medical Microbiology, University Hospital, Groningen, The Netherlands
Rights & Permissions [Opens in a new window]

Summary

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the 'Save PDF' action button.

A mobile laminar cross-flow unit for surgery has been evaluated by the use of an experimental bacterial aerosol of Escherichia coli in different concentrations, generated at several different sites. A separate ventilation system, mounted underneath the table, produced an almost downward directed curtain of sterile air along both sides of the table. When the velocity of the air stream, discharged by the cross-flow unit, was adjusted at 0·50 m./sec. at 2 m. from the filter face (at the head end of the operation table), the surface of the table could be maintained free of bacterial settling, even after aerosolization of heavily concentrated suspensions of 108 bacteria/ml. at different sites outside the enclosure and underneath the table as well.