Hostname: page-component-7b7567c8c-fvjsw Total loading time: 0 Render date: 2026-11-17T07:31:53.635Z Has data issue: false hasContentIssue false

Description and analysis of a simple micro-titration immune cytolytic test

Published online by Cambridge University Press:  15 May 2009

R. L. Spooner
Affiliation:
Department of Pathology, University of Cambridge
F. W. Bowden
Affiliation:
Department of Human Ecology, University of Cambridge
R. G. Carpenter
Affiliation:
Department of Human Ecology, University of Cambridge
Rights & Permissions [Opens in a new window]

Extract

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the 'Save PDF' action button.

A method of micro-titration for cytolytic antibodies is described. Standard drops of diluent are placed on a subdivided siliconed slide and dilutions are made by rotating a wire loop holding 1 drop of serum in each drop sequentially. One drop of a mixture of cells and complement is then added. The slide is incubated in a moist chamber for 90 min at 37° C. after which a small cover-slip is placed on each drop and the amount of cell lysis, judged by phase-contrast microscopy, is graded.

An analysis of variance of the method has been performed and the method has been found to be reliable, accurate and unbiased.

It allows accurate estimates of cytolytic activity in sera using cell concentrations of 104 to 105 cells per ml. Eight ten-dilution titrations can be carried out with 1 ml. of such a suspension.

Information

Type
Research Article
Copyright
Copyright © Cambridge University Press 1965