Autologous total ossicular replacement prosthesis

Published online by Cambridge University Press:  11 November 2014

M Malhotra*
Affiliation:
Department of Otorhinolaryngology, All India Institute of Medical Sciences, Rishikesh, Uttarakhand, India
S Varshney
Affiliation:
Department of Otorhinolaryngology, All India Institute of Medical Sciences, Rishikesh, Uttarakhand, India
R Malhotra
Affiliation:
Department of Anatomy, All India Institute of Medical Sciences, Rishikesh, Uttarakhand, India
*
Address for correspondence: Dr Manu Malhotra, 3/4, Type IV, AIIMS, Rishikesh 249201, Uttarakhand, India E-mail: manumalhotrallrm@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Objective:

To develop an autologous total ossicular replacement prosthesis with sustainable hearing results.

Methods:

The ears of 40 patients, who had chronic otitis media with absent suprastructure of the stapes and long process of the incus, were repaired using the autologous total ossicular replacement technique. Post-operative results were evaluated after 6 and 12 months on the basis of average pure tone air conduction and average air–bone gap measured at 0.5, 1, 2 and 3 kHz.

Results:

Successful rehabilitation of pure tone average to 30 dB or less was achieved in 75 per cent of patients, and air–bone gap to 20 dB or less was attained in 82.5 per cent of patients. Overall mean improvement in air–bone gap was 23.9 ± 8.5 dB (p < 0.001). Mean improvements in air–bone gap were significantly greater (p < 0.05) in the tympanoplasty only group (27.3 ± 6.6 dB) and the intact canal wall tympanoplasty group (25.9 ± 6.3 dB) than in the canal wall down tympanoplasty group (16.3 ± 8.9 dB).

Conclusion:

This paper describes an autologous total ossicular replacement prosthesis that is biocompatible, stable, magnetic resonance imaging compatible and, above all, results in sustainable hearing improvement.

Information

Type
Main Articles
Copyright
Copyright © JLO (1984) Limited 2014 
Figure 0

Table IPatient and surgical data

Figure 1

Fig. 1The autologous total ossicular replacement prosthesis (TORP) surgical procedure entails: (a) exposure of conchal cartilage (arrow) via post-aural incision, (b) drilling of a hole in conchal cartilage and (c) reshaping of the bone with a drill. The image in part (d) shows a vertical section of the TORP resting on the stapes footplate.

Figure 2

Table IIOverall ABG and PTA pre- and post-operative, and improvement data

Figure 3

Fig. 2Overall post-operative air–bone gap findings.

Figure 4

Table IIIPost-operative ABG and PTA data

Figure 5

Table IVPost-operative ABG and PTA data by procedure type

Figure 6

Table VABG and PTA hearing gains by procedure type

Figure 7

Table VISummary of hearing results

Figure 8

Table VIIcomparison of ABG and PTA results with previous literature on other prostheses