A High Throughput 14C Accelerator Mass Spectrometer

Published online by Cambridge University Press:  18 July 2016

Kenneth H. Purser*
Affiliation:
Southern Cross Corporation, 426C Boston Street, Topsfield, Massachusetts 01983 USA
Rights & Permissions [Opens in a new window]

Abstract

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the 'Save PDF' action button.

I present design details of a tandem accelerator mass spectrometer, which has been installed at the National Ocean Sciences AMS Facility at Woods Hole, Massachusetts, to provide precision 14C/13C/12C isotopic ratios for sub-milligram-size samples of graphite with throughputs of >4000 samples per year. A unique feature is the capability for simultaneous measurement of all three isotopes after acceleration, to avoid differential transmission effects and to allow on-line fractionation corrections and diagnosis of instrument health. Using filamentous graphite fabricated from a recent sample, we have established the counting rate of 14C ions at between 60–120 s–1.