Low-Level Gas Proportional Counting in An Underground Laboratory

Published online by Cambridge University Press:  18 July 2016

H H Loosli
Affiliation:
Physics Institute, University of Bern, Sidlerstrasse 5, CH-3012 Bern, Switzerland
Martin Heimann
Affiliation:
Physics Institute, University of Bern, Sidlerstrasse 5, CH-3012 Bern, Switzerland
Hans Oeschger
Affiliation:
Physics Institute, University of Bern, Sidlerstrasse 5, CH-3012 Bern, Switzerland
Rights & Permissions [Opens in a new window]

Abstract

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the 'Save PDF' action button.

Results of measurements in an underground laboratory, 70m of water equivalent below surface, are given. Background values of proportional gas counters with volumes between 16cc and 1.5L are lower by a factor of 2—4.5 compared to the values in the previously used laboratory (7m w.e.). High counting gas pressures, up to 5 at CH4 and up to 36 at P-10, enable the use of relatively small counters with correspondingly small background contributions from the walls. A separation of the residual background into different components is attempted, distinguishing a pressure dependent volume effect and surface correlated contributions. It can be shown that the selection of radioactively pure counter construction material is very important for a good low level counting system.