Hostname: page-component-7b7567c8c-2n9c9 Total loading time: 0 Render date: 2026-11-16T21:54:41.334Z Has data issue: false hasContentIssue false

Electro-Convulsive Therapy with Minimum Hazard

Published online by Cambridge University Press:  29 January 2018

Douglas Gordon*
Affiliation:
The City University, St John Street, London EC1V 4PB and Honorary Consultant Radiologist, Moorfields Eye Hospital, London
Rights & Permissions [Opens in a new window]

Summary

Core share and HTML view are not available for this content. However, as you have access to this content, a full PDF is available via the 'Save PDF' action button.

The inter-electrode resistance during ECT is shown to be very variable, a finding which invalidates the measurement of shocks in joules. With square wave pulses of selected current value the threshold for convulsion may be below 100 milliamperes or above 2,000 milliamperes. It is advisable to measure the current value during millisecond pulses with relatively long intervals between.

Information

Type
Research Article
Copyright
Copyright © The Royal College of Psychiatrists 

This journal is not currently accepting new eletters.

eLetters

No eLetters have been published for this article.