Subglacial measurement of turbidity and electrical conductivity

Published online by Cambridge University Press:  20 January 2017

Dan B. Stone
Affiliation:
Department of Geophysics and Astronomy, University of British Columbia, Vancouver, British Columbia V6T 1Z4, Canada
Garry K. C. Clarke
Affiliation:
Department of Geophysics and Astronomy, University of British Columbia, Vancouver, British Columbia V6T 1Z4, Canada
Erik W. Blake
Affiliation:
Department of Geophysics and Astronomy, University of British Columbia, Vancouver, British Columbia V6T 1Z4, Canada
Rights & Permissions [Opens in a new window]

Abstract

Direct measurements of the properties of subglacial water are necessary for understanding water flow beneath glaciers. In this paper we describe the construction, calibration and field usage of two instruments—one that measures turbidity and the other that measures electrical conductivity of subglacial water. The sensors are inexpensive and reliable. To demonstrate the potential usefulness of these devices, we present samples of data obtained from beneath Trapridge Glacier, Yukon Territory, Canada.

Information

Type
Instruments and Methods
Copyright
Copyright © International Glaciological Society 1993
Figure 0

Fig. 1. Turbidity sensor schematic and circuit diagram showing the relative positions and physical connections of electronic components, a. The basic components are an incandescent light source, two infrared photo-Darlington detectors – one for the reference light path and one for the light path through the water sample–and a voltage regulator. These components are housed in a plastic shell and then sealed in a casting resin that does not degrade in the presence of mineralized water, b. Circuit diagram for turbidity sensor and voltage measurement. Shown in this diagram are the unregulated supply voltage from the surface (upper left), the sensor components (middle and right) and the surface-voltage measurement circuit (lower left).

Figure 1

Fig. 2. Electrical-conductivity sensor showing the basic design and measurement circuit. Stainless steel rods are press-fitted into a pre-drilled, solid nylon cylinder. To allow solder connections for lead wires, brass pins are inserted into the interior ends of the rods. The entire assembly is placed in a section of ABS or PVC conduit, which is then filled with casting resin. With this device, the conductance of basal water can be determined by means of an a.c. half-bridge measurement.

Figure 2

Fig. 3. Observations of subglacial turbidity from different sensors beneath Trapridge Glacier, Yukon Territory, during three consecutive field seasons: 1989 (a), 1990 (b) and 1991 (c). Day 188 corresponds to 7 July. These data illustrate the wide range of responses that we have measured with our sensors–from small-amplitude diurnal fluctuation (a), to large-amplitude turbidity events (b). Both resbonse types are evident in (c).

Figure 3

Fig. 4. Electrical conductivity of water beneath Trapridge Glacier measured with different sensors in 1990 (a) and 1991 (b). Day 188 corresponds to 7 July. The data shown in (a) were obtained from a sensor that was installed during the 1989 field season. Note that in (a) the large conductivity decrease on day 204 coincides with the turbidity pulse shown in Figure 3b.