Hostname: page-component-74656f9884-rdgjk Total loading time: 0 Render date: 2026-11-07T11:42:30.355Z Has data issue: false hasContentIssue false

14C MICROSAMPLE ANALYSIS WITH ECHoMICADAS FACILITIES: CURRENT STATE OF PLAY

Published online by Cambridge University Press:  20 September 2024

François Thil*
Affiliation:
Laboratoire des Sciences du Climat et de l’Environnement, IPSL/LSCE, CEA CNRS UVSQ, Université Paris-Saclay, F-91191, Gif Sur Yvette, France
Nadine Tisnérat-Laborde
Affiliation:
Laboratoire des Sciences du Climat et de l’Environnement, IPSL/LSCE, CEA CNRS UVSQ, Université Paris-Saclay, F-91191, Gif Sur Yvette, France
Christine Hatté
Affiliation:
Laboratoire des Sciences du Climat et de l’Environnement, IPSL/LSCE, CEA CNRS UVSQ, Université Paris-Saclay, F-91191, Gif Sur Yvette, France Division of Geochronology and Environmental Sciences, Institute of Physics, Silesian University of Technology, 44-100 Gliwice, Poland
Elias Kader
Affiliation:
Laboratoire des Sciences du Climat et de l’Environnement, IPSL/LSCE, CEA CNRS UVSQ, Université Paris-Saclay, F-91191, Gif Sur Yvette, France
Claude Noury
Affiliation:
Laboratoire des Sciences du Climat et de l’Environnement, IPSL/LSCE, CEA CNRS UVSQ, Université Paris-Saclay, F-91191, Gif Sur Yvette, France
Martine Paterne
Affiliation:
Laboratoire des Sciences du Climat et de l’Environnement, IPSL/LSCE, CEA CNRS UVSQ, Université Paris-Saclay, F-91191, Gif Sur Yvette, France
Brian Phouybanhdyt
Affiliation:
Laboratoire des Sciences du Climat et de l’Environnement, IPSL/LSCE, CEA CNRS UVSQ, Université Paris-Saclay, F-91191, Gif Sur Yvette, France
Lukas Wacker
Affiliation:
Ion Beam Physics, Department of Physics, ETH Zürich, HPK H29, Otto-Stern-Weg 5, 8093 Zürich, Switzerland
*
*Corresponding author. Email: francois.thil@lsce.ipsl.fr
Rights & Permissions [Opens in a new window]

Abstract

In 2015, a new accelerator mass spectrometry facility (AMS), the ECHoMICADAS (Environnement, Climat, Homme, MIni CArbon DAting System), was installed in the Laboratoire des Sciences du Climat et de l’Environnement (LSCE). Equipped with a hybrid source, it allows the analysis of solid or gas samples for 14C measurement. Here, we summarize the equipment surrounding the Gas Interface System (GIS), namely the elemental analyzer (EA), the carbonate handling system (CHS2) and the ampoule cracker. We describe our model of sample contamination, taking into account the cross and the constant contaminations, and then describe how these contaminations were handled in the data processing. Both contaminant corrections are applied before the phases of blank subtraction and standard normalization, making it possible to use the standard, blank and sample ratios without contaminant during these phases. We finally present our results on normalization standards (N=118), blanks (N=125) and reference materials (N=117) for different measurement protocols and for sample masses between 3 and 300 µgC.

Information

Type
Conference Paper
Copyright
© The Author(s), 2024. Published by Cambridge University Press on behalf of University of Arizona
Figure 0

Figure 1 ECHoMICADAS GIS equipment. Three introduction possibilities: Elemental analyzer (EA), Carbonate Handling System (CHS2), and ampoule cracker system. In this configuration, the syringe is connected to the AMS while the CO2 from the EA or the CHS2 is fixed on the cold zeolite trap (22ºC). Once the syringe is empty, it is connected to the zeolite trap; the trap, which is heated to 450ºC, releases the CO2 in the syringe.

Figure 1

Figure 2 (a) Constant and cross contamination model; (b) Impact of the data reduction on two different 14C/12C ratios: Oxa2 and blank. Steps 1, 3, 4 and 5 are the same calculation steps as the ones applied in BATS software and are described in Wacker et al. (2010). Step 2 is added to take into account the mass of the sample. Step 1: molecular current 13CH correction. Step 2: contaminant correction based on the formula (1) described in Figure 2a. Step 3: blank correction Step 4: mass fractionation correction. Step 5: ratio normalization

Figure 2

Figure 3 Standard and blank F14C values versus carbon mass, between 3 and 300 µgC (log scale). Only LSCE 2021 and 2022 measurements with high energy 12C current higher than 3 µA are reported. Figure 3a—Organic matter standards and blanks measured on EA-GIS without step 2 (black squares), with step 2 (gray squares) and with the constant contaminant correction step done at the end (blue squares), Figure 3b—Carbonate standards (IAEA-C1 and IAEA-C2) and in-house standard (Patella Vulgata) measured on the CHS2-GIS (black and gray squares for ratio without or with step 2 correction) and measured via ampoule cracker-GIS (blue and red circles for ratios without or with step 2 correction).

Figure 3

Table 1 F14C values for the international standards oxalic acid II, IAEA-C2, IAEA-C7, IAEA-C8 and IAEA-C1 and phthalic acid (PHA) blanks. Blank values are reported without contaminant correction and blank subtraction. Three different means were calculated for masses higher than 100 µgC, medium masses between 35 and 100 µgC and masses lower than 35 µgC. The number of measurements N is given in brackets.

Figure 4

Figure 4 IAEA-C2 and IAEA-C1 F14C values versus elapsed time, between CO2 sample tube sealing on the BCA line and cracker-GIS measurement on ECHoMICADAS. IAEA-C1 values are not corrected from blank and contaminant.