Effects of resolution in real and reciprocal spaces from a 2D detector at a high-energy synchrotron beamline

Published online by Cambridge University Press:  02 January 2018

Andrea Bernasconi*
Affiliation:
Università di Pavia, Pavia, Italy
Jonathan Wright
Affiliation:
ESRF – The European Synchrotron, Grenoble, France
*
a) Author to whom correspondence should be addressed. Electronic mail: andrea.bernasconi@unipv.it

Abstract

Different experimental conditions at a versatile high-energy beamline equipped with a two-dimensional detector have been compared for powder diffraction and pair distribution function (PDF) experiments. In particular, sample size and sample to detector distances have been evaluated on a standard sample, to evaluate their effects in both Q and real space. Two illustrative cases are also discussed. The average structure and local distortions in a BaTiO3 powder with 100 nm particle size show that spurious ripples in the PDF are suppressed by increased counting statistics. Effects of small amounts of a crystalline impurity phase on a SiO2.Al2O3.Na2O.CaO glass have been quantified.

Information

Type
Technical Articles
Copyright
Copyright © International Centre for Diffraction Data 2018 
Figure 0

Table I. Quartz wt. % addition to the glass phase.

Figure 1

Figure 1. Geometry of the diffraction experiment in transmission with a 2D detector.

Figure 2

Figure 2. (Colour online) Computed effects of different parameters on deltaQ as defined from Eq. (6). Initial configuration (i.e. blue curve) corresponds to a sample thickness of 0.5 mm, a beamsize of 75 μm, a point spread of 4 pixels (assuming a pixel size of 50 μm, a bandpass of 2*10−3, a wavelength of 0.1899 Å, and a sample to detector distance of 9 cm. Variations from this initial configuration are described in the different legends.

Figure 3

Figure 3. (Colour online) Measured FWHMs of the six investigated configurations versus Q (symbols) and deltaQ/2 from Eq. (6) (lines).

Figure 4

Figure 4. (Colour online) Silicon NIST SRM 640c G(r)s for different Qrange and capillary size. On the left the real space resolution is displayed. On the right side the real space dampening is displayed. For blue curve the QMAX is about 24 Å−1 while for green curves QMAX is about 20 Å−1.

Figure 5

Figure 5. (Colour online) PDFGUI silicon NIST SRM 640c G(r) graphic fit for all the different investigated sample to detector distances and capillary size conditions. QDAMP and QBROAD are also displayed.

Figure 6

Figure 6. (Colour online) Experimental coordination numbers of silicon NIST SRM 640c obtained by fitting the radial distribution function R(r).

Figure 7

Table II. Average percentage error in silicon coordinations as determined by radial distribution function analysis.

Figure 8

Figure 7. (Colour online) Graphical fit of BaTiO3 Rietveld refinement. HKLs are provided for BaTiO3 (in green) and for BaCO3 (in grey).

Figure 9

Table III. BaTiO3 results from Rietveld refinement of the tetragonal and cubic crystallographic models at room temperature.

Figure 10

Figure 8. (Colour online) PDF data of the investigated BaTiO3. Top left: raw data (sample in blue, empty capillary in green). Top right: reduced total scattering function Q.F(Q) for 1 single vs 500 images with a magnified inset for the high Q region. Bottom left: G(r) in the 0–80 Å range. Bottom right: G(r) is plotted in the local 1–7.2 Å range for 1 single vs 500 images.

Figure 11

Table IV. BaTiO3 pair distribution function results in the average region (15–45 Å range).

Figure 12

Table V. Pair distribution function results in the local region (1.5–15 Å range).

Figure 13

Figure 9. 2θ vs intensity pattern of glass samples with different quartz amounts. On the left side, as collected pattern are displayed while on the right side the median filtering has been applied to reduce Bragg signal from quartz.