Echocardiographic versus angiographic measurement of the aortic valve annulus in children undergoing balloon aortic valvuloplasty: method affects outcomes

Published online by Cambridge University Press:  14 October 2020

George T. Nicholson*
Affiliation:
Vanderbilt University Medical Center, Vanderbilt University School of Medicine, Nashville, TN, USA
Bryan H. Goldstein
Affiliation:
The Heart Institute, Department of Pediatrics, University of Cincinnati School of Medicine, Cincinnati Children’s Hospital Medical Center, Cincinnati, OH, USA
Kevin Gao
Affiliation:
Sibley Heart Center, Department of Pediatrics, Emory University School of Medicine, Children’s Healthcare of Atlanta, Atlanta, GA, USA
Ritu Sachdeva
Affiliation:
Sibley Heart Center, Department of Pediatrics, Emory University School of Medicine, Children’s Healthcare of Atlanta, Atlanta, GA, USA
Sean M. Lang
Affiliation:
The Heart Institute, Department of Pediatrics, University of Cincinnati School of Medicine, Cincinnati Children’s Hospital Medical Center, Cincinnati, OH, USA
Scott Gillespie
Affiliation:
Department of Biostatistics, Emory University School of Medicine, Atlanta, GA, USA
Sung-in H. Kim
Affiliation:
Sibley Heart Center, Department of Pediatrics, Emory University School of Medicine, Children’s Healthcare of Atlanta, Atlanta, GA, USA
Christopher J. Petit
Affiliation:
Sibley Heart Center, Department of Pediatrics, Emory University School of Medicine, Children’s Healthcare of Atlanta, Atlanta, GA, USA
*
Author for correspondence: George T. Nicholson, MD, Assistant Professor of Pediatrics, Vanderbilt University School of Medicine, 220 Children’s Way, Nashville, TN 37232, USA. Tel: +1 615 322 7447; Fax: +1 615 322 2210. E-mail: george.t.nicholson@vumc.org
Rights & Permissions [Opens in a new window]

Abstract

Objective:

Operators are mindful of the balloon-to-aortic annulus ratio when performing balloon aortic valvuloplasty. The method of measurement of the aortic valve annulus has not been standardised.

Methods and results:

Patients who underwent aortic valvuloplasty at two paediatric centres between 2007 and 2014 were included. The valve annulus measured by echocardiography and angiography was used to calculate the balloon-to-aortic annulus ratio and measurements were compared. The primary endpoint was an increase in aortic insufficiency by ≥2 degrees. Ninety-eight patients with a median age at valvuloplasty of 2.1 months (Interquartile range (IQR): 0.2–105.5) were included. The angiographic-based annulus was 8.2 mm (IQR: 6.8–16.0), which was greater than echocardiogram-based annulus of 7.5 mm (IQR: 6.1–14.8) (p < 0.001). This corresponded to a significantly lower angiographic balloon-to-aortic annulus ratio of 0.9 (IQR: 0.9–1.0), compared to an echocardiographic ratio of 1.1 (IQR: 1.0–1.1) (p < 0.001). The degree of discrepancy in measured diameter increased with smaller valve diameters (p = 0.041) and in neonates (p = 0.044). There was significant disagreement between angiographic and echocardiographic balloon-to-aortic annulus ratio measures regarding “High” ratio of >1.2, with angiographic ratio flagging only 2/12 (16.7%) of patients flagged by echocardiographic ratio as “High” (p = 0.012). Patients who had an increase in the degree of aortic insufficiency post valvuloplasty, only 3 (5.5%) had angiographic ratio > 1.1, while 21 (38%) had echocardiographic ratio >1.1 (p < 0.001). Patients with resultant ≥ moderate insufficiency more often had an echocardiographic ratio of >1.1 than angiographic ratio of >1.1 There was no association between increase in balloon-to-aortic annulus ratio and gradient reduction.

Conclusions:

Angiographic measurement is associated with a greater measured aortic valve annulus and the development of aortic insufficiency. Operators should use caution when relying solely on angiographic measurement when performing balloon aortic valvuloplasty.

Information

Type
Original Article
Copyright
© The Author(s), 2020. Published by Cambridge University Press
Figure 0

Figure 1. Measurement of the aortic valve annulus diameter in the parasternal long-axis view during echocardiography. The aortic valve annulus diameter was measured as the distance between the hinge points for the visualised anterior and posterior leaflets.

Figure 1

Table 1. Demographic and catheterisation characteristics.

Figure 2

Table 2. Paired differences in balloon-to-annulus ratio by imaging modality.

Figure 3

Table 3. Demographic and clinical characteristics by successive BAR disagreement percent in aortic valve annulus diameter.

Figure 4

Figure 2. Repeated measures two-factor one-way analysis of variance interaction mean value plots considering balloon-to-aortic annulus ratio by imaging modality and age groups. Results indicate significant differences in balloon-to-aortic annulus ratio values between imaging modalities for all age categories. This difference appears to be most profound in neonates. Estimates are presented as mean values and 95% confidence intervals.

Figure 5

Table 4. Differences in BAR measurements between moderate AI groups.

Figure 6

Figure 3. Balloon-to-aortic annulus ratio and change in aortic valve insufficiency by measurement method. Scatterplots illustrating balloon-to-aortic annulus ratio versus change in aortic valve insufficiency as measured by imaging modalities. More cases of severe changes in aortic valve insufficiency were captured by balloon-to-aortic annulus ratio > 1.0 with balloon-to-aortic annulus ratio determined by echocardiogram (plot on the right). Almost all cases where aortic valve insufficiency changed by 2 more had an echocardiographic balloon-to-aortic annulus ratio > 1.0.

Figure 7

Figure 4. Analysis of covariance plot considering the relationship between catheter reduction in gradient and balloon-to-aortic annulus ratio by imaging modality. analysis of covariance analysis demonstrating reduction in gradient (mmHg) against balloon-to-aortic annulus ratio. Filled dots represent angiographic balloon-to-aortic annulus ratio, and hollow dots represent echocardiographic balloon-to-aortic annulus ratio. The line with the more positive slope (dashed line) represents the line of best fit for echocardiography. No significance difference in slopes was observed (p = 0.494).