Improvised bespoke technique for atrial septostomy in the shortage of atrioseptostomy balloon catheters

Published online by Cambridge University Press:  25 August 2023

Raymond N. Haddad*
Affiliation:
M3C-Necker, Hôpital Universitaire Necker-Enfants malades, Assistance Publique-Hôpitaux de Paris (AP-HP), Paris, France
Zainab Alawadhi
Affiliation:
Department of Pediatric Cardiology, Heart Centre of Excellence, Al Jalila Children’s Speciality Hospital, Dubai, United Arab Emirates
Mahmoud Al Soufi
Affiliation:
Department of Pediatric Cardiology, Heart Centre of Excellence, Al Jalila Children’s Speciality Hospital, Dubai, United Arab Emirates
Mohamed Kasem
Affiliation:
Department of Pediatric Cardiology, Heart Centre of Excellence, Al Jalila Children’s Speciality Hospital, Dubai, United Arab Emirates
*
Corresponding author: R. N. Haddad; Email: raymondhaddad@live.com
Rights & Permissions [Opens in a new window]

Abstract

Atrioseptostomy balloon catheter is an essential item to have on our shelves. However, the recall and shortage in production of the commonly used balloon atrioseptostomy catheters posed an imminent threat to our patients. Herein, we present the case of a newborn with a post-natal diagnosis of simple transposition of great arteries and restrictive atrial communication where repeated static balloon atrial septostomy using a 9 mm x 20 mm Sterling balloon failed to improve his status. We had to improvise per-operatively a new bespoke technique to perform a vital pull-through balloon atrial septostomy. The distal third of a 10 mm x 20 mm semi-compliant Cristal balloon was exteriorised out the tip of a 6-Fr 55 cm Cook Flexor sheath in the left atrium, and both were simultaneously pulled back to the right atrium to create an 8 mm septal defect. The procedure was successful without any complications. The baby was weaned off prostaglandin on day 3 and surgically repaired on day 5 with excellent results.

Information

Type
Brief Report
Copyright
© The Author(s), 2023. Published by Cambridge University Press
Figure 0

Figure 1. Bench-side preparation and testing of the improvised bespoke balloon atrial septostomy using a 10 mm x 20 mm Cristal balloon (white pointed arrow) and a 6-Fr 55 cm Cook Flexor sheath (black pointed arrow).

Figure 1

Figure 2. Anteroposterior (A) and lateral (B) fluoroscopic views of improvised bespoke balloon atrial septostomy positioned into the left atrium. Note the two 0.014-inch Balance Middleweight wires in the lower left and upper right pulmonary veins (white stars) over which the distal third of a 10 mm x 20 mm semi-compliant Cristal balloon (white pointed arrow) is exteriorised out the tip of a 6-Fr 55 cm Cook Flexor sheath (black pointed arrow).

Supplementary material: File

Haddad et al. supplementary material 1
Download undefined(File)
File 13.1 MB
Supplementary material: File

Haddad et al. supplementary material 2
Download undefined(File)
File 862.6 KB
Supplementary material: File

Haddad et al. supplementary material 3
Download undefined(File)
File 23.5 MB