Dosimetric properties of fluoroscopic EPID for transit dosimetry

Published online by Cambridge University Press:  30 October 2014

Akbar Anvari*
Affiliation:
Department of Radiation Medicine Engineering, Shahid Beheshti University, Tehran, Iran
Seyed Mahmoud Reza Aghamiri
Affiliation:
Department of Radiation Medicine Engineering, Shahid Beheshti University, Tehran, Iran
Seyed Rabi Mahdavi
Affiliation:
Department of Medical Physics, Iran University of Medical Sciences, Tehran, Iran
Parham Alaei
Affiliation:
Department of Radiation Oncology, University of Minnesota, Minneapolis, Minnesota, USA
Mohammad Mohammadi
Affiliation:
Department of Medical Physics, Royal Adelaide Hospital, Adelaide, SA, Australia
*
Correspondence to: Akbar Anvari, Department of Radiation Medicine Engineering, Shahid Beheshti University, Tehran 19839-63113, Iran. Tel: +98 9196260656. E-mail: A_Anvari@sbu.ac.ir
Rights & Permissions [Opens in a new window]

Abstract

The aim of this work was to evaluate dose response of fluoroscopic EPID for transit dosimetry applications. Properties studied included warm up time, build-up thickness evaluation, dose history, linearity, stability, and short and long-term reproducibility of EPID response, as well as field size dependence.

Pixel value matrices of electronic portal images in DICOM format were analysed in central and 8 off axis points using customised written codes in Matlab. In order to do this, nine 26×26 pixel matrices were selected as regions of interest, the regions represented by these arrays were 1×1 and 0·65×0·65 cm2 at the EPID and isocentre level, respectively.

Necessary warm up time for stable operation of EPID is 30 minutes, and there is no need for extra build-up layer to increase the dose response. Linearity tests indicate charged coupled device camera of EPID saturates at 50 cGy level, and does not have linear relationship with dose. Reproducibility and stability of the measurements were excellent and the detector showed same signal with a maximum deviation of <0·3% both in short and long terms. Results of dosimetric evaluation have shown the TheraView fluoroscopic EPID can be used for transit dosimetry purposes.

Information

Type
Original Articles
Copyright
© Cambridge University Press 2014 
Figure 0

Figure 1 The position of the nine regions of interest (ROI) selected for investigation of EPI physical characteristics, field size 10×10 cm2 at isocentre level.

Figure 1

Figure 2 The linearity test for the electronic portal imaging devices (EPID) in 2×2 and 10×10 cm2 areas of central region, the measurements are shown as symbols, and results of the nonlinear regression (solid line) are fitted to them. Monitor units used were: 2, 3, 5, 10, 20, 30, 40, 50, 60, 70, 80, 90, 100 and 200.

Figure 2

Figure 3 The linearity test for the electronic portal imaging devices (EPID) in the center 2×2 cm2 area of central regions of interest (ROI) for field size of 10×10 cm2. The measurements are shown as symbols, and results of the polynomial 4 regression (solid line) are fitted to them. Monitor units used were: 2, 5, 10, 20, 30, 40 and 50.

Figure 3

Figure 4 The variation of the electronic portal imaging devices (EPID) pixel values to the absolute dose values. The data for central point and eight peripheral points described in Figure 1 were acquired; each data point is the average of three consecutive measurements.

Figure 4

Figure 5 Dose profiles of charged coupled device (CCD) camera-based electronic portal imaging devices (EPID) for open field in a 10×10 cm2 field at different monitor units (MUs): 2, 5, 10, 20, 30, 40 and 50.

Figure 5

Figure 6 The mean pixel values of short-term and long-term reproducibility test for central point and eight peripheral points against number of use.

Figure 6

Figure 7 The variation of minimum, mean and maximum pixel values of consecutive electronic portal imaging devices (EPID) images on the 2×2 cm2 area of central regions of interest (ROI) as a function of number of use.

Figure 7

Figure 8 The field size response for square field sizes varies from 2×2 and 18×18 cm2. Central regions of interest (ROI) measurements are shown as symbols; results of the cubic regression (solid line) are also shown.

Figure 8

Figure 9 Relative open field profiles of electronic portal imaging devices (EPID) for various field sizes delivered by 20 MU.