Do blood contamination reduction devices work? A single institution comparison

Published online by Cambridge University Press:  25 November 2024

Maria E. Navas*
Affiliation:
Louis Stokes VA Medical Center, VA Northeast Ohio Healthcare System, Cleveland, OH, USA
Salman Siddiq
Affiliation:
Louis Stokes VA Medical Center, VA Northeast Ohio Healthcare System, Cleveland, OH, USA
Laurie Bauer
Affiliation:
Louis Stokes VA Medical Center, VA Northeast Ohio Healthcare System, Cleveland, OH, USA
Jose A. Rivera
Affiliation:
Louis Stokes VA Medical Center, VA Northeast Ohio Healthcare System, Cleveland, OH, USA
Anita J. White
Affiliation:
Louis Stokes VA Medical Center, VA Northeast Ohio Healthcare System, Cleveland, OH, USA
Stella Ache
Affiliation:
Louis Stokes VA Medical Center, VA Northeast Ohio Healthcare System, Cleveland, OH, USA
Mark Osborne
Affiliation:
Louis Stokes VA Medical Center, VA Northeast Ohio Healthcare System, Cleveland, OH, USA
Nataliya Kachaluba
Affiliation:
Louis Stokes VA Medical Center, VA Northeast Ohio Healthcare System, Cleveland, OH, USA
Brian Klonowski
Affiliation:
Louis Stokes VA Medical Center, VA Northeast Ohio Healthcare System, Cleveland, OH, USA
Christine Robbins
Affiliation:
Louis Stokes VA Medical Center, VA Northeast Ohio Healthcare System, Cleveland, OH, USA
Curtis Donskey
Affiliation:
Louis Stokes VA Medical Center, VA Northeast Ohio Healthcare System, Cleveland, OH, USA
*
Corresponding author: Maria E. Navas; Email: marunavas@yahoo.com
Rights & Permissions [Opens in a new window]

Abstract

We compare two initial specimen diversion devices evaluated over 3 months to investigate their utility in lowering blood culture contamination rates at or below 1%. Overall contamination rates during trial periods were 2.46% and 2.60% but usage was low, whereas device-specific contamination rates were 0.68% and 0.8%, respectively.

Information

Type
Concise Communication
Copyright
© The Author(s), 2024. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America
Figure 0

Figure 1.Comparison of overall contamination rates during trial and the year prior.

Figure 1

Table 1.Blood culture contamination rates during trial