A study on modelling cochlear duct mid-scalar length based on high-resolution computed tomography, and its effect on peri-modiolar and mid-scalar implant selection

Published online by Cambridge University Press:  19 August 2019

G Pamuk
Affiliation:
Department of Otorhinolaryngology, Kırıkkale Yüksek İhtisas Hospital, Kırıkkale, Turkey
A E Pamuk*
Affiliation:
Department of Otorhinolaryngology, Kırıkkale Yüksek İhtisas Hospital, Kırıkkale, Turkey
A Akgöz
Affiliation:
Department of Radiology, Faculty of Medicine, Hacettepe University, Ankara, Turkey
E Öztürk
Affiliation:
Department of Biostatistics, Faculty of Medicine, Hacettepe University, Ankara, Turkey
M D Bajin
Affiliation:
Department of Otorhinolaryngology, Faculty of Medicine, Hacettepe University, Ankara, Turkey
L Sennaroğlu
Affiliation:
Department of Otorhinolaryngology, Faculty of Medicine, Hacettepe University, Ankara, Turkey
*
Author for correspondence: Dr Ahmet Erim Pamuk, Department of Otorhinolaryngology, Kırıkkale Yüksek İhtisas Hospital, Kırıkkale 71400, Turkey E-mail: dr_erim@hotmail.com Fax: +90 318 215 1195
Rights & Permissions [Opens in a new window]

Abstract

Objective

To determine cochlear duct mid-scalar length in normal cochleae and its role in selecting the correct peri-modiolar and mid-scalar implant length.

Methods

The study included 40 patients with chronic otitis media who underwent high-resolution computed tomography of the temporal bone. The length and height of the basal turn, mid-modiolar height of the cochlea, mid-scalar and lateral wall length of the cochlear duct, and the ‘X’ line (the largest distance from mid-point of the round window to the mid-scalar point of the cochlear canal) were measured.

Results

Cochlear duct lateral wall length (28.88 mm) was higher than cochlear duct mid-scalar length (20.08 mm) (p < 0.001). The simple linear regression equation for estimating complete cochlear duct length was: cochlear duct length = 0.2 + 2.85 × X line.

Conclusion

Using the mid-scalar point as the reference point (rather than the lateral wall) for measuring cochlear duct mid-scalar length, when deciding on the length of mid-scalar or peri-modiolar electrode, increases measurement accuracy. Mean cochlear duct mid-scalar length was compatible with peri-modiolar and mid-scalar implant lengths. The measurement method described herein may be useful for pre-operative peri-modiolar or mid-scalar implant selection.

Information

Type
Main Articles
Copyright
Copyright © JLO (1984) Limited, 2019 
Figure 0

Fig. 1. Axial computed tomography image demonstrating measurements of: (a) basal turn length, (b) basal turn height and (c) mid-modiolar cochlea height. The sections were passed through the round window. A = anterior; P = posterior

Figure 1

Fig. 2. Cochlear duct length based on three-dimensional multi-planar reconstruction: (a) marked cochlear duct and (b) straightened cochlear duct.

Figure 2

Fig. 3. A coronal reformatted computed tomography section showing the ‘X’ line, defined as the largest distance between the mid-point of round window and the mid-scalar point of the cochlear canal. HAL = head-anterior-left; RAF = right-anterior-feet; LPH = left-posterior-head; FPR = feet-posterior-right

Figure 3

Table 1. Descriptive statistics of five measurements

Figure 4

Fig. 4. Boxplots of five measurements with respect to gender and ear side.

Figure 5

Fig. 5. Scatter plot of the ‘X’ line and cochlear duct mid-scalar length. CDL = cochlear duct length