L-band radiometer design for long-term stability measurement of one-port devices

Published online by Cambridge University Press:  14 December 2011

Emilie Leynia de la Jarrige*
Affiliation:
LAAS CNRS, 7 avenue du colonel Roche, 31077 Toulouse, Cedex, France. Université de Toulouse; UPS, INSA, INP, ISAE; LAAS; F-31077 Toulouse, France.
Laurent Escotte
Affiliation:
LAAS CNRS, 7 avenue du colonel Roche, 31077 Toulouse, Cedex, France. Université de Toulouse; UPS, INSA, INP, ISAE; LAAS; F-31077 Toulouse, France.
Eric Gonneau
Affiliation:
SIMAD, Université de Toulouse, F-31077 Toulouse, France.
Jean-Marc Goutoule
Affiliation:
EADS Astrium, 31402 Toulouse, France.
*
Corresponding author: E. Leynia de la Jarrige Email: eleyniad@laas.fr
Rights & Permissions [Opens in a new window]

Abstract

The design of a high-stability radiometer is presented in this paper. It is used to characterize the stability of a microwave active cold load at L-band. The two-load radiometer takes advantage of a noise injection technique to improve its sensitivity. A temperature-stabilized enclosure is used to minimize gain and noise temperature fluctuations of the receiver. The radiometer sensitivity is 0.031 K and the gain fluctuations are less than 0.005 dB during 1 day. The total gain and the receiver noise temperature of the radiometer exhibit very small variations (<0.03 dB and 1 K, respectively) on the long term.

Information

Type
Research Papers
Copyright
Copyright © Cambridge University Press and the European Microwave Association 2011
Figure 0

Fig. 1. Block-diagram of the experimental setup. The red marks on the microwave circuits correspond to the thermistors.

Figure 1

Fig. 2. Simplified block diagram of the two-load radiometer.

Figure 2

Fig. 3. Typical measurement cycle.

Figure 3

Fig. 4. Photograph of the temperature-stabilized system.

Figure 4

Fig. 5. Simplified block-diagram of the temperature control circuit.

Figure 5

Fig. 6. ACL physical temperature versus time recorded from October 1, 2010 to February 15, 2011. The errors bars correspond to ±σ.

Figure 6

Table 1. Temperature of microwave circuits inside the radiometer.

Figure 7

Fig. 7. Allan variance versus integration time. The solid lines correspond to the model given by (18).

Figure 8

Table 2. Radiometer characteristics determined from Allan variance analysis.

Figure 9

Fig. 8. Total gain of the radiometer versus time recorded from October 1, 2010 to February 15, 2011. The errors bars correspond to ±σ. The vertical dotted lines indicate the period where the normalized gain fluctuation spectrum plotted in Fig. 9 is calculated.

Figure 10

Fig. 9. Normalized gain fluctuation of the radiometer versus frequency. The solid line corresponds to the model given by (19).