Hostname: page-component-74656f9884-2dmjx Total loading time: 0 Render date: 2026-11-07T02:44:41.702Z Has data issue: false hasContentIssue false

An experimental study of the vertical travel accuracy of the treatment table used in radiotherapy

Published online by Cambridge University Press:  09 August 2012

Hans L. Riis*
Affiliation:
Radiofysisk Laboratorium, Odense University Hospital, Odense C, Denmark
Sune J. Zimmermann
Affiliation:
Radiofysisk Laboratorium, Odense University Hospital, Odense C, Denmark
*
Correspondence to: Hans L. Riis, Radiofysisk Laboratorium, Odense University Hospital, Sdr. Boulevard 29, DK–5000 Odense C, Denmark. Tel: + 45 29 79 64 73. E-mail: hans.lynggaard.riis@ouh.regionsyddanmark.dk
Rights & Permissions [Opens in a new window]

Abstract

The idea of the present work was to investigate general features of the vertical travel of the treatment table in different configurations and the limitations of an optimised alignment. The investigations were carried out on two tables for different load cases, lateral positions and turntable angles. A wire was held vertically nearby the rotational axis of the table. The wire was used as reference in the investigations. A digital USB-microscope was attached to the tabletop. An orthogonal set of images of the wire was acquired at different vertical table positions. By analysing the images the vertical travel accuracy of the table was extracted. The two tables were found to travel linearly with the same characteristics over the full range of 110 cm with deviations less than 0.5 mm relative to a straight line. A divergence from a vertical travel above 0.5 mm was found to originate from misalignment. An alignment procedure to attain the optimal vertical performance with a minimum of inclination was presented. This procedure must be relayed on objective measured data of the travel with uncertainties far beyond 0.5 mm. Our suggested experimental method was found to have the potential to obtain the required data.

Information

Type
Original Article
Copyright
Copyright © Cambridge University Press 2012
Figure 0

Figure 1. Schematic diagram showing the treatment table and the reference system defined by the wire. The plumb bob is seen immersed into the liquid in the glass. The tabletop is shown at isocentre height z = 0.0 cm approximately 125 cm above floor level. The longitudinal position of the tabletop was 30.0 cm. The wire was positioned within a few millimetres from the table rotational axis. The distance between the wire and the superior end of the tabletop was 7 cm. A force F was applied 38 cm from superior end of the tabletop. The figure is based on Figure 14 at doi: 10.1088/0031-9155/55/24/014. Reproduction was permitted by IOP Publishing Ltd.

Figure 1

Figure 2. Lifting mechanism (light-blue) and tabletop (gray) seen from above. The iso-rotation arm is not displayed. The quantity E denotes the eccentric rotational axis. The force F was acting in the longitudinal and vertical symmetry plane of the tabletop. The distance between the z- and E-axes is 100 cm. (a) The definition of the coordinate systems (x, y, z), (x′, y′, z′) and the turntable angle θ appear from the figure. The y-axis is towards the gantry. The tabletop is shown at the lateral position 0.0 cm. (b) Three different lateral positions of the tabletop are shown: (i) –25.0 cm, (ii) 0.0 cm and (iii) 25.0 cm.

Figure 2

Figure 3. Iso-rotation base frame seen from above. The gray part (base frame) is fixed to the foundation with double M20 nuts at each four positions denoted A2, A1, B1 and B2. Levelling of the unit is performed by adjusting the nuts. According to the Elekta Precise Table Installation Manual, the A2 and B2 nuts are tighten to 160 Nm while A1 and B1 are tighten to 10 Nm. The edge C0 is standing on the bottom of the pit. The orange part is called the iso-rotation arm. This part is also shown in Figure 1. The rotation of the arm is limited to θ = [−100°; 100°]. The lengths are w1 ≅ 200 mm, w2 ≅ 550 mm and ℓ ≅ 578 mm.

Figure 3

Figure 4. Measured data of the tabletop movements relative to the wire at different table heights Δz. The turntable angle was held at θ = 0° and no external loads were applied. The red and blue curves denote Ac1 and Ac4 data, respectively. The markers\scale50%\squareand\scale50%\xcircdenote Δx′ and Δy′ values, respectively. (a) The data represent upward tabletop movements only. The black curves are polynomial fit of first and second degree. All displayed data were measured on 14 February, 2011. (b) Data for both upwards and downwards movements of the tabletop are shown. The displayed data were measured on 7 October, 2008 (Ac1) and 2 July, 2008 (Ac4).

Figure 4

Figure 5. Tabletop movements relative to the wire at different table heights Δz. The displayed data were measured on Ac4 on 2 July, 2008. Curves representing 0 kg (green), 40 kg (blue) and 80 kg (red) are shown at the lateral table positions −25.0 cm (\triangleleft), 0.0 cm (\scale40%\xcirc), and 25.0 cm (\triangleright). (a) The x′-direction. (b) The y′-direction.

Figure 5

Figure 6. The data are displayed as Δx′ (magenta) and Δy′ (cyan) at turntable angles θ = −90° (\triangleright), θ = 0° (\triangle) and θ = 90° (\triangleleft). The external load was 80 kg and the lateral position was 0.0 cm. For θ ≠ 0°, position data were measured at only three heights. (a) Tabletop movements measured at Ac4 on 2 July, 2008. (b) The results after optimisation of the data in Figure 6(a) by using ψopt=−0.029° and Φopt=0.003° in (7).