Aspects of simulator cone-beam CT for radiotherapy treatment planning

Published online by Cambridge University Press:  15 September 2010

Denise Irvine*
Affiliation:
Regional Medical Physics Service, Belfast HSC Trust, Belfast, UK
Mark McJury
Affiliation:
Regional Medical Physics Service, Belfast HSC Trust, Belfast, UK
*
Correspondence to: Denise Irvine, Regional Medical Physics Service, Belfast HSC Trust, Lisburn Road, Belfast BT9 7AB, UK. E-mail: denise.irvine@belfasttrust.hscni.net
Rights & Permissions [Opens in a new window]

Abstract

Background and purpose: Following a recent major upgrade in cone-beam computed tomography (CBCT) software and functionality, we have reassessed aspects of our Varian Acuity simulator performance for use in treatment planning. The feasibility of using CBCT for treatment planning has been assessed and here we report specifically on Hounsfield number (HN) accuracy and related dose errors, and digitally reconstructed radiograph (DRR) image quality.

Methods: Using a Catphan® 600 CT phantom, HN accuracy and uniformity were investigated for a range of CBCT imaging modes. This included the variation in HNs with scan length and phantom position. Results were compared with those acquired from conventional CT. Treatment plans for three sites were generated using the Rando phantom, and results from CBCT-based data were compared to that from CT-based data using a gamma analysis. Image quality of DRRs based on CBCT data were compared with those from CT data both quantitatively, by calculating the modulation transfer function (MTF) and qualitatively, by counting the number of line pairs visible on a phantom.

Results and conclusions: Catphan data showed that for certain cases, the HN calibration of the Acuity CBCT was out of tolerance and could lead to errors in dose calculation of >2%. HNs were only acceptable for scan lengths >10 cm. In multi-scan mode, geometric shifts and differences in HNs were seen on CT slices on either side of the interface between the two acquisitions. However, comparisons between treatment plans calculated using CBCT data and conventional CT data from Rando phantoms showed that head, pelvis and thorax plans were acceptable. CBCT DRR image quality compared favourably with a conventional CT scanner in some respects; however, image uniformity and low contrast resolution were poorer due to the ‘cupping’ artefact obtained with CBCT scans.

Information

Type
Original Article
Copyright
Copyright © Cambridge University Press 2010
Figure 0

Table 1. Planning parameters and details of the CBCT acquisitions for the Rando phantom scans

Figure 1

Table 2. HN values and tolerances derived from the results from the Siemens scanner, the Varian recommended specifications and results obtained using a range of imaging protocols in a conventional CT scanner

Figure 2

Table 3. Hounsfield number variation with decreasing scan length for a full-fan scan, standard dose 150 cm

Figure 3

Figure 1. Reconstructed anterior–posterior image of the Catphan CT phantom scanned using the double cone-beam computed tomography mode, with the ‘step’ artefact arrowed.

Figure 4

Table 4. HNs measured for ‘double’ CBCT scans

Figure 5

Figure 2. HN uniformity: (a) full-fan scan; (b) half-fan scan; (c) plot of Hounsfield number (HN) against distance across phantom.

Figure 6

Figure 3. Comparison of isodose lines generated using cone-beam computed tomography and computed tomography data for the Rando phantom pelvis.

Figure 7

Figure 4. Calculated modulation transfer function (MTF) from bar pattern DRRs. std res, standard resolution; high res, high resolution.

Figure 8

Table 5. The frequency for which the modulation transfer function falls to 50% of its value is shown for full and half-fan CBCT imaging modes and for two slice thicknesses (2.5 mm and 5.0 mm)

Figure 9

Figure 5. DRRs created from CT data (left) and CBCT data (right). The CBCT DRRs are both geometrically accurate and adequate for the purposes of matching to portal images.