Hostname: page-component-74656f9884-2dmjx Total loading time: 0 Render date: 2026-11-10T04:20:01.874Z Has data issue: false hasContentIssue false

Development and dosimetric evaluation of the IMRT prostate at outside-the-irradiated field in a heterogeneity male pelvis phantom

Published online by Cambridge University Press:  29 June 2022

J. Jayamani
Affiliation:
School of Health Sciences, Health Campus, Universiti Sains Malaysia, 16150 Kubang Kerian, Kelantan, Malaysia Oncological and Radiological Sciences Cluster, Advanced Medical and Dental Institute, Universiti Sains Malaysia, 13200 Kepala Batas, Pulau Pinang, Malaysia
N. D. Osman
Affiliation:
Oncological and Radiological Sciences Cluster, Advanced Medical and Dental Institute, Universiti Sains Malaysia, 13200 Kepala Batas, Pulau Pinang, Malaysia
A. A. Tajuddin
Affiliation:
Albukhary International University, 05200 Alor Setar, Kedah, Malaysia
D. O. Samson
Affiliation:
Department of Physics, University of Abuja, P.M.B 117, F.C.T., Abuja, Nigeria
K. E. Kamaruddin
Affiliation:
School of Health Sciences, Health Campus, Universiti Sains Malaysia, 16150 Kubang Kerian, Kelantan, Malaysia
M. Z. Abdul Aziz*
Affiliation:
Oncological and Radiological Sciences Cluster, Advanced Medical and Dental Institute, Universiti Sains Malaysia, 13200 Kepala Batas, Pulau Pinang, Malaysia
*
Author for correspondence: Mohd Zahri Abdul Aziz, Oncological and Radiological Sciences Cluster, Advanced Medical and Dental Institute, Universiti Sains Malaysia, 13200 Kepala Batas, Pulau Pinang, Malaysia. Email: mrzahri@gmail.com/mohdzahri@usm.my
Rights & Permissions [Opens in a new window]

Abstract

Background:

Intensity-modulated radiation therapy (IMRT) treatment delivery requires pre-treatment patient-specific quality assurance (QA) for the dosimetry verification due to its complex multileaf-collimator movement. The prostate target close position between the bladder and rectum requires a tight margin during planning, and mistreatment would have a huge impact on the patient. A commercially available QA tool consists of a homogeneous medium and does not represent an exact photon interaction on the tumour and also on the nearby healthy organ.

Objective:

A heterogeneous male pelvis phantom was developed and investigated the efficiency of the treatment planning system (TPS) calculation on the off-axis region.

Methods:

Polymethyl methacrylate was used for the phantom housing, and the material closed to the bladder, rectum and prostate density was chosen to construct the organ models. The phantom was scanned and validated by the computed tomography number and density. An IMRT treatment was planned in the Monaco TPS, and a thermoluminescent dosimeter (TLD-100) was used to validate the point dosimetry. In addition, an EGSnrc Monte Carlo simulation was carried out to validate the phantom dosimetry.

Results & Discussion:

The dose measurement between TLD-100, TPS, and EGSnrc was compared and validated in the pelvis phantom. In the prostate region, the dose difference was within ± 5%, and the maximum dose difference outside-the-irradiated field was up to 20·07 % and 47·31 % in TPS and TLD-100, respectively. Meanwhile, the measured dose was lower than the calculated dose, and it was apparent for the dose outside-the-irradiated field.

Conclusion:

The developed heterogeneity male pelvis phantom was validated and verified to be an important QA device for validating radiation dosimetry in the pelvis region. The dose outside-the-irradiated field was underestimated by both TPS and TLD, respectively.

Information

Type
Original Article
Copyright
© The Author(s), 2022. Published by Cambridge University Press
Figure 0

Table 1. The density of human tissue and the material selected for male pelvis phantom construction. Measured value using Archimedes principle compared with the NIST X-COM value

Figure 1

Figure 1. (a) Octagon phantom housing with measurement 33 cm × 22 cm × 30 cm, and (b) organ models were inserted in the phantom housing, (c) the arrangement of the bladder, prostate and rectum in the pelvis phantom.

Figure 2

Figure 2. CT DICOM image (trans-axial view) at the isocenter point of the heterogeneity phantom in MONACO TPS in CT DICOM format. The white dot on the rectum shows the TLD-100 chip and the rod within the models is the PMMA rod that hold the organ models together.

Figure 3

Table 2. Physical density and electron density (ED) of the materials used to model the male pelvis phantom. The percentage difference of the ED between the patient and the phantom material. Mean CT number of the phantom materials and the patient CT number. The patient data and CT number were retrieved from a (Shrotriya et al., 2018)

Figure 4

Figure 3. The transaxial view of TPS dose distribution displayed on the (a) Monaco TPS and (b) EGSnrc.

Figure 5

Figure 4. The measured dose (TLD) and calculated dose (TPS and EGSnrc) for the selected points in (a) prostate, (c) bladder and (e) rectum in the phantom. The relative dose difference for TPS and EGSnrc was calculated relative to measure dose (TLD) in (b) prostate, (d) bladder and (f) rectum.

Supplementary material: File

Jayamani et al. supplementary material

Jayamani et al. supplementary material

Download Jayamani et al. supplementary material(File)
File 2.2 MB