Semi-Automated Equipment for CO2 Purification and Graphitization at the A.E. Lalonde AMS Laboratory (Ottawa, Canada)

Published online by Cambridge University Press:  17 August 2016

Gilles St-Jean
Affiliation:
Department of Earth and Environmental Sciences and A.E. Lalonde Lab, University of Ottawa, 25 Templeton St., Ottawa, Ontario, K1N 6N5, Canada
William E Kieser
Affiliation:
Department of Physics and A.E. Lalonde Lab, University of Ottawa, 150 Louis Pasteur, Ottawa, Ontario, K1N 6N5, Canada
Carley A Crann*
Affiliation:
Department of Earth and Environmental Sciences and A.E. Lalonde Lab, University of Ottawa, 25 Templeton St., Ottawa, Ontario, K1N 6N5, Canada
Sarah Murseli
Affiliation:
Department of Earth and Environmental Sciences and A.E. Lalonde Lab, University of Ottawa, 25 Templeton St., Ottawa, Ontario, K1N 6N5, Canada
*
*Corresponding author. Email: radiocarbon@uottawa.ca.
Rights & Permissions [Opens in a new window]

Abstract

New computer-controlled, semi-automatic systems were designed and built for CO2 purification and graphitization at the A.E. Lalonde Accelerator Mass Spectrometry (AMS) Laboratory with consideration for user friendliness and high throughput. The stainless steel vacuum lines are orbitally welded to ensure clean seams with low memory. The insulated graphitization ovens with plug-in electrodes provide a hazard-free environment for operators. The closed-loop cooling system circulating low-viscosity Dynalene at –40°C provides highly efficient water trapping. The LabVIEWTM software features (1) pressure and temperature recording for QA/QC; (2) safety interlocks to preclude operator errors resulting in sample loss, cross-contamination, or damaging a vacuum pump; and (3) automation for leak checking, iron conditioning, and running samples. Results from the first year of routinely measured standards, reference, and background materials are reproducible and within acceptance values. In the first year of operation (commissioned in spring 2014), over 1000 targets (~60% unknowns) were produced. With new tube sealing and CO2 purification lines, and two more graphitization lines now operational, the Lalonde AMS Laboratory is able to provide routine radiocarbon analysis (>200 µg carbon) at a capacity of more than 7000 targets per year. Most importantly, the equipment is safe and intuitive, making it ideal for education and training students to run their own samples.

Information

Type
Advances in Physical Measurement Techniques
Copyright
© 2016 by the Arizona Board of Regents on behalf of the University of Arizona 
Figure 0

Table 1 Summary of common equipment design components described in the text.

Figure 1

Figure 1 Comparison of stainless steel welds: (a) smooth and complete weld by an orbital welder; (b) a nonpenetrating hand weld; and (c) an overheated hand weld.

Figure 2

Figure 2 Main LabVIEW interface used to control the graphitization lines. The configuration presented here shows the reaction cells being filled with hydrogen through the manifold during the automated iron conditioning routine. Once the reaction cells have reached 700 mbar, the pump valves will close, the hydrogen gas is shut off, and the heaters (currently showing residual heat from the oxidation step) will ramp up to 520°C. At this stage, the Cool Temp is reading room temperature since the circulator pump on the cooling system is not yet turned on.

Figure 3

Figure 3 (a) Photograph of the graphitization line with an oven (1) removed to show the quartz reaction tube with Fe powder (2), the water trap (3), and pressure transducer (4). (5) Rotulex cracker assembly (glass joint tube with Cu turnings at the bottom to cushion the movement of the breakseal); (6) cooling cups for water removal; (7) touch-screen monitor (LabVIEW controls) mounted on rollers above the electronics box (8), and (9) the second display showing the monitoring software for pressure (top), cooling bar temperature (middle), and heater temperature (bottom). (b) and (c) show a closer look at the cracker and reactor assemblies, respectively (numbers the same as in a).

Figure 4

Figure 4 (a) Cryogenic system for water removal with the front cover open and part of the insulaton removed to show the feed lines to the 10 cooling cups. The labeled parts are as follows: (1) linear track actuators to raise and lower the cold trap table (2); (3) cooling cups (inset showing copper coil); (4) chiller containing Dynalene at –40°C; (5) circulator; (6) circulator speed control. (b) Phase diagram for water: the vertical gray bar denotes the temperature and range in pressure conditions experienced by the H2O produced during the graphitization reaction.

Figure 5

Figure 5 Monitoring software for the graphitization line. The main phases of graphitization are labeled on the plot of the pressure transducers (top), reading left to right: Fe oxidation (O2); pumping; Fe reduction (H2); pumping; sample release (CO2); addition of H2; graphitization. The middle plot shows the temperature of each cooling cup, and the bottom plot shows the temperature of each heater. Each plot shows the time in a different manner: start to finish in minutes (top); start time using 24-hr clock (middle); start time using AM/PM (bottom).

Figure 6

Figure 6 Automation tab on the display, generated by the LabVIEW software. In this scenario, the iron conditioning routine is at the hydrogen cleaning step.

Figure 7

Figure 7 CO2 purification line: (1) stainless steel Rotulex crackers containing samples in quartz breakseal tubes; (2) cleaning trap, shown here filled with liquid nitrogen, but usually filled with –80°C ethanol slurry; (3) quantitative trap consisting of a Dewar on a linear track actuator and pressure transducers; (4) Pyrex breakseals for clean CO2; (5) vacuum tube and arrow pointing toward HiCube pumping station under the table; (6) monitor stand (on wheels connected to 80/20 frame) and arrow pointing to touch-screen monitor with LabVIEW software (not shown).

Figure 8

Table 2 Results for blank and reference materials tested during the first year of operation. References: IAEA (Rozanski et al. 1990; Rozanski 1991); AVR-07-PAL-37 (Reyes et al. 2010); Hollis Mine mammoth [constrained by ~700-ka glass-fission track age on tephra from Westgate et al. (2011), and Valley Creek tephra from Jensen et al. (2013)].

Figure 9

Figure 8 Memory test results for one of the graphitization lines at the University of Ottawa (referred to as G2 in the lab).