The impact of the inverse chirp z-transform on breast microwave radar image reconstruction

Published online by Cambridge University Press:  28 April 2020

Tyson Reimer*
Affiliation:
Department of Physics and Astronomy, University of Manitoba, Winnipeg, MB, Canada
Mario Solis-Nepote
Affiliation:
Research Institute in Oncology and Hematology, University of Manitoba, Winnipeg, MB, Canada
Stephen Pistorius
Affiliation:
Department of Physics and Astronomy, University of Manitoba, Winnipeg, MB, Canada Research Institute in Oncology and Hematology, University of Manitoba, Winnipeg, MB, Canada
*
Author for correspondence: Tyson Reimer, E-mail: reimert5@myumanitoba.ca
Rights & Permissions [Opens in a new window]

Abstract

This work examines the impact of the inverse chirp z-transform (ICZT) for frequency-to-time-domain conversion during image reconstruction of a pre-clinical radar-based breast microwave imaging system operating over 1–8 GHz. Two anthropomorphic breast phantoms were scanned with this system, and the delay-multiply-and-sum beamformer was used to reconstruct images of the phantoms, after using either the ICZT or the inverse discrete Fourier transform (IDFT) for frequency-to-time domain conversion. The contrast, localization error, and presence of artifacts in the reconstructions were compared. The use of the IDFT resulted in prominent ring artifacts that were not present when using the ICZT, and the use of the ICZT resulted in higher contrast between the tumor and clutter responses. In one of the phantoms, the tumor response was only visible in reconstructions that used the ICZT. The use of the ICZT evaluated with a time-step size of 11 ps resulted in the reduction of prominent artifacts present when using the IDFT and the successful identification of the tumor response in the reconstructed images.

Information

Type
Research Paper
Copyright
Copyright © Cambridge University Press and the European Microwave Association 2020
Figure 0

Fig. 1. 3D-printed MRI-based breast phantoms used in this investigation. Fibroglandular (left) and adipose (right) shells used to create a Class III phantom in (a), and a Class I phantom in (b).

Figure 1

Fig. 2. TD radar signals obtained using the IDFT and ICZT (evaluated at 1024 points between 0 and 6 ns response times). Signals obtained from one antenna position from an experimental scan of the Class III breast phantom.

Figure 2

Fig. 3. DMAS reconstructions of a 15 mm radius lesion in a Class I phantom using (a) the IDFT and (b) the ICZT. The solid green circle indicates the tumor location, the dotted white circle indicates the approximate phantom boundary, and the dotted black circle indicates the antenna trajectory during the scan. Each image is normalized to its maximum intensity value.

Figure 3

Fig. 4. DMAS reconstructions of a 15 mm radius lesion in a Class III phantom using (a) the IDFT and (b) the ICZT. The solid green circle indicates the tumor location, the dotted white circle indicates the approximate phantom boundary, and the dotted black circle indicates the antenna trajectory during the scan. Each image is normalized to its maximum intensity value.

Figure 4

Fig. 5. SMR of the DMAS reconstructions produced after using either the IDFT or the ICZT (evaluated at 1024 time-points between 0 and 6 ns response times).

Figure 5

Fig. 6. SCR of the DMAS reconstructions produced after using either the IDFT or the ICZT (evaluated at 1024 time-points between 0 and 6 ns response times).

Figure 6

Fig. 7. Localization error of the Class III reconstructions as a function of the time-step size. The dotted line indicates the radius of the tumor.

Figure 7

Fig. 8. DMAS reconstructions of a 15 mm radius lesion in a Class III phantom using (a) the IDFT, (b) the ICZT with 100 time-points from 0 to 6 ns (60.6 ps time-step size), (c) the ICZT with 200 time-points (30.2 ps time-step size), and (d) the ICZT with 550 time-points (10.9 ps time-step size).

Figure 8

Fig. 9. Average total absolute-difference of images produced using time-step sizes greater than 5.9 ps and the image produced after using the ICZT for FD-to-TD conversion with a time-step size of 5.9 ps, for 10 scans of the Class III phantom. Standard deviations of the average total absolute-difference over the images from the 10 scans are the displayed uncertainties. The shaded red region indicates the 99.7% confidence intervals on the fit.