Monte Carlo study on mucosal dose in oral and naval cavity using photon beams with small field

Published online by Cambridge University Press:  11 January 2011

James C.L. Chow*
Affiliation:
Department of Radiation Physics, Princess Margaret Hospital, Toronto, Ontario, Canada Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada Department of Physics, Ryerson University, Toronto, Ontario, Canada
Amir M. Owrangi
Affiliation:
Department of Radiation Physics, Princess Margaret Hospital, Toronto, Ontario, Canada
*
Correspondence to: James C.L. Chow, Department of Radiation Physics, Princess Margaret Hospital, Toronto, Ontario, M5G 2M9 Canada. E-mail: james.chow@rmp.uhn.on.ca
Rights & Permissions [Opens in a new window]

Abstract

We study how mucosal dose in the oral or nasal cavity depends on the irradiated small segmental photon fields varying with beam energy, beam angle and mucosa thickness. Dose ratio (mucosal dose with bone underneath to dose at the same point without bone) reflecting the dose enhancement due to the bone backscatter was determined by Monte Carlo simulation (EGSnrc-based code), validated by measurements. Phase space files based on the 6 and 18 MV photon beams with small field size of 1 × 1 cm2, produced by a Varian 21 EX linear accelerator, were generated using the BEAMnrc Monte Carlo code. Mucosa phantoms (mucosa thickness = 1, 2 and 3 mm) with and without a bone under the mucosa were irradiated by photon beams with gantry angles varying from 0 to 30°. Doses along the central beam axis in the mucosa and the dose ratio were calculated with different mucosa thicknesses. For the 6 MV photon beams, the dose at the mucosa-bone interface increased by 44.9–41.7%, when the mucosa thickness increased from 1 to 3 mm for the beam angle ranging from 0 to 30°. These values were lower than those (58.8–53.6%) for the 18 MV photon beams with the same beam angle range. For both the 6 and 18 MV photon beams, depth doses in the mucosa were found to increase with an increase of the beam angle. Moreover, the dose gradient in the mucosa was greater for the 18 MV photon beams compared to the 6 MV. For the dose ratio, it was found that the dose enhancement due to the bone backscatter increased with a decrease of mucosa thickness, and was more significant at both the air-mucosa and mucosa-bone interface. Mucosal dose with bone was investigated by Monte Carlo simulations with different experimental configurations, and was found vary with the beam energy, beam angle and mucosa thickness for a small segmental photon field. The dosimetric information in this study should be considered when searching for an optimized treatment strategy to minimize the mucosal complications in the head-and-neck intensity-modulated radiation therapy.

Information

Type
Original Article
Copyright
Copyright © Cambridge University Press 2010
Figure 0

Figure 1. Schematic diagram (not to scale) showing the beam geometry and experimental configuration used in Monte Carlo simulations. 6 and 18 MV photon beams with field size of 1 × 1 cm2 were used. The thickness of the bone slab was equal to 5 cm.

Figure 1

Figure 2. (a,b) The percentage depth doses and beam profiles for the 6 MV photon beams with field size of 1 × 1 cm2, respectively. (c,d) The percentage depth doses and beam profiles for the 18 MV photon beams. Doses were measured and calculated by photon diode and Monte Carlo simulation with SSD = 90 cm. The beam profiles were determined at a depth of 10 cm.

Figure 2

Figure 3. Depth doses in the mucosa with thicknesses of (a) 1, (b) 2 and (c) 3 mm varying with beam angles of 0, 10, 20 and 30° using the 6 MV photon beams.

Figure 3

Figure 4. Depth doses in the mucosa with thicknesses of (a) 1, (b) 2 and (c) 3 mm varying with beam angles of 0, 10, 20 and 30° using the 18 MV photon beams.

Figure 4

Figure 5. Relationship between dose and beam angle at different depths in the mucosa with thicknesses of (a) 1, (b) 2 and (c) 3 mm using the 6 MV photon beams.

Figure 5

Table 1. Tables showing dose ratios of mucosa with to without bone for the 6 MV photon beams varying with depth and beam angle with mucosa thicknesses of (a) 1, (b) 2 and (c) 3 mm

Figure 6

Table 2. Tables showing dose ratios of mucosa with to without bone for the 18 MV photon beams varying with depth and beam angle with mucosa thicknesses of (a) 1, (b) 2 and (c) 3 mm

Figure 7

Figure 6. Relationship between dose and beam angle at different depths in the mucosa with thicknesses of (a) 1, (b) 2 and (c) 3 mm using the 18 MV photon beams.