Effects of density from various hip prosthesis materials on 6 MV photon beam: a Monte Carlo study

Published online by Cambridge University Press:  21 February 2017

M. Z. Abdul Aziz*
Affiliation:
Oncological and Radiological Science Cluster, Advanced Medical and Dental Institute, Universiti Sains Malaysia, Penang, Malaysia
F. N. Mohd Kamarulzaman
Affiliation:
School of Physics, Universiti Sains Malaysia, Penang, Malaysia
N. A. S. Mohd Termizi
Affiliation:
School of Physics, Universiti Sains Malaysia, Penang, Malaysia
N. Abdul Raof
Affiliation:
Oncological and Radiological Science Cluster, Advanced Medical and Dental Institute, Universiti Sains Malaysia, Penang, Malaysia
A. A. Tajuddin
Affiliation:
School of Physics, Universiti Sains Malaysia, Penang, Malaysia
*
Correspondence to: M. Z. Abdul Aziz, Oncological and Radiological Science Cluster, Advanced Medical and Dental Institute, Universiti Sains Malaysia, 13200 Kepala Batas, Penang, Malaysia. E-mail: mr.zahri@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

In radiotherapy planning, computed tomography (CT) images are used to calculate the dose in the patient. However, a high density hip prosthesis can cause streaking artefacts in CT images, which make dose calculations for nearby organs inaccurate. This study aim to quantify the impact of a hip prosthesis on 6 MV photon beam dose distribution using the Monte Carlo (MC) simulation. To quantify the radiation dose at the hip prosthesis accurately, image processing techniques were used to generate CT images free from streak artefacts. MATLAB software was used to produce computer-generated phantoms consisting of bone, titanium, stainless steel and CoCrMo. Percentage depth dose (PDD) and beam profile were used to analyse the impact of the hip prosthesis on the dose distribution of the photon beam. PDD showed that the absorbed dose was reduced as the density of the material increased, and the dose was reduced by as much as 49% when the photon beam struck the highest density material (CoCrMo, 8·2g/cm3). However, dose was increased at the tissue-hip prosthesis interface (depths of 4 and 19cm). As the depth increased, the absorbed dose decreased due to attenuation of photons by the tissue and the metal.

Information

Type
Original Articles
Copyright
© Cambridge University Press 2017 
Figure 0

Figure 1 (a) A phantom made from Perspex and Teflon used in this study. (b) An image of the phantom scanned using a 64 multislice computed tomography simulator system. Notes: Teflon can be observed clearly in the image.

Figure 1

Table 1 Density and composition of materials used in the DOSXYZnrc simulation

Figure 2

Figure 2 Photons with energy 6 MV were directed lateral to the phantom. Notes: The first and second Teflon pieces are located 4 and 19 cm, respectively, from the surface.

Figure 3

Figure 3 Spectrum of 6 MV photons from the International Atomic Energy Agency phase-space database (Siemens Primus).

Figure 4

Figure 4 Scatter plot analysis illustrates a precise 10×10 cm2 field size of the simulated model.

Figure 5

Figure 5 Percentage depth dose of the 6 MV photon beam oriented transverse to the various materials with different densities [water, bone, titanium (Ti), stainless steel (SST) and CoCrMo]. Notes: Drastic perturbation of doses were observed at depths of 4–9 cm (first femoral head position) and 19–24 cm (second femoral head position). Water ρ, 1 g/cm3; bone ρ, 2·08 g/cm3; Ti ρ, 4·48 g/cm3; SST ρ, 6·45 g/cm3; and CoCrMo ρ, 8·20 g/cm3.

Figure 6

Figure 6 Percentage difference of dose perturbation of the 6 MV photon beam oriented transverse to various materials with different densities [water, bone, titanium (Ti), stainless steel (SST) and CoCrMo]. Notes: Bone, titanium and stainless steel effects on dose distribution were as high as 7, 32 and 41%, respectively, compared with water. The highest perturbation was observed for CoCrMo, for which a 49% difference was calculated.