Dosimetry evaluation of 192Ir Flexisource using gafchromic EBT3 film

Published online by Cambridge University Press:  06 November 2024

Aziz Oustous*
Affiliation:
Physical Sciences and Engineering, ENPT, Moulay Ismail University, Faculty of Sciences, Zitoune, Meknès, Morocco Regional Oncology Center, Beni Mellal, Morocco
Zouhair Airouss
Affiliation:
Laboratory of Engineering in Chemistry and Physics of Matter (LICPM), Sultan Moulay Slimane University, Faculty of Sciences and Techniques, Beni Mellal, Morocco
Yacine EL Attaoui
Affiliation:
Laboratory of Engineering in Chemistry and Physics of Matter (LICPM), Sultan Moulay Slimane University, Faculty of Sciences and Techniques, Beni Mellal, Morocco
Yassine Oustous
Affiliation:
Doctoral Study Center for Life and Health Sciences, Faculty of Medicine and Pharmacy of Rabat, Impasse Souissi, Rabat, Morocco
Mohammed Khalis
Affiliation:
Physical Sciences and Engineering, ENPT, Moulay Ismail University, Faculty of Sciences, Zitoune, Meknès, Morocco
Rajaa Sebihi
Affiliation:
Department of Physics, Mohamed V University, Faculty of Sciences, Rabat, Morocco
*
Corresponding author: Aziz Oustous; Email: aziz.oustous@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Introduction:

Brachytherapy is a widely used cancer treatment technique requiring precise dose delivery for optimal outcomes. This study evaluates the dosimetric properties of the ${{\rm{\;}}^{192}}$Ir Flexisource using Gafchromic EBT3 film, adhering to American Association of Physicists in Medicine guidelines.

Methods:

The EBT3 film was calibrated with a linear accelerator to establish the relationship between optical density and absorbed dose. Measurements were conducted in a water-equivalent solid phantom to simulate human tissue. The dosimetric properties measured included the dose rate constant, radial dose function and anisotropy function.

Results:

The results showed a close alignment with reference data: a maximum deviation of 3·1% for the dose rate constant, 2·3% for the radial dose function and 5·67% for the anisotropy function. These deviations are within acceptable limits, indicating high accuracy.

Conclusions:

Gafchromic EBT3 film demonstrates reliable dosimetric measurement in high-dose rate brachytherapy. The film’s accuracy in evaluating the ${{\rm{\;}}^{192}}$Ir Flexisource ensures precise treatment delivery and enhances patient safety, supporting its continued use in clinical practice.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Figure 1. Geometry of the Ir-192 Flexisource used in this study.

Figure 1

Figure 2. Scanned images of calibration and background films.

Figure 2

Figure 3. The dose-response curve of EBT3.

Figure 3

Figure 4. Setup for exposing EBT3 radiochromic films (RCFs) using PMMA plates.

Figure 4

Figure 5. Dose distribution around the 192Ir Flexisource.

Figure 5

Figure 6. Differences between radial dose functions calculated with EBT3 film and Granero et al.9

Figure 6

Figure 7. Anisotropy function at distances from 1 to 4 cm and angles from 0° to 180°: EBT3 film versus Granero et al.9