Atrial flow regulator in transposition of the great arteries, ventricular septal defect, and pulmonary stenosis: a case report

Published online by Cambridge University Press:  02 October 2024

Saïd Bichali*
Affiliation:
Univ. Lille, CHU Lille, Paediatric Cardiology, Lille, France
Alexandre Delarue
Affiliation:
Univ. Lille, CHU Lille, Paediatric Cardiology, Lille, France
Ali Houeijeh
Affiliation:
Univ. Lille, CHU Lille, Paediatric Cardiology, Lille, France
*
Corresponding author: Saïd Bichali; Email: said_b91@hotmail.fr
Rights & Permissions [Opens in a new window]

Abstract

A 5-month-old girl with transposition of the great arteries, ventricular septal defect, and pulmonary stenosis underwent the implantation of a 6-mm M Atrial Flow Regulator (Occlutech) over an 8-French delivery sheath for significant cyanosis and progressive restriction of the atrial septal defect, without adverse event for 6 months. The Atrial Flow Regulator device could improve blood mixing as a bridge to surgery.

Information

Type
Brief Report
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Figure 1. Heart ultrasound scans showing an unrestricted left-to-right atrial shunt after atrioseptostomy at birth (a), a restricted left-to-right atrial shunt at 5 months of age (b) with increased Doppler maximal velocity to 1.8 m/s (c), a left-to-right shunt and seldom right-to-left shunt on the day following the implantation of a 6-mm M Atrial Flow Regulator (AFR, Occlutech) on a bicaval view (d), the AFR device barely touching the aortic wall on a parasternal short axis view (e), a reasonable left-to-right Doppler maximal velocity of 1.5 m/s (f), and a colour diameter of the left-to-right shunt across the AFR of 5.4 mm on a 4-chamber subcostal view (g).

Figure 1

Figure 2. Ninety-degree right anterior oblique angiography revealing a millimetric rounded arterial duct (a), anteroposterior angiography showing the patent stented arterial duct with left-to-right-shunt from the aorta to the pulmonary arteries at the time of stent implantation at 1.5 months of age (b), 10-degree left anterior oblique projection with subtraction showing a stent fracture at the aortic end with separation of stent fragments (c), anteroposterior view after ductal stent-in-stent and Atrial Flow Regulator (AFR, Occlutech) device implantation at 5 months of age (d), en-face chest X-ray at 9 months of age revealing a recurrent stent fracture (e), and 10-degree cranial projection with a right coronary catheter crossing the patent AFR on a hydrophilic-coated (Terumo) guidewire with 2-mmHg trans-atrial gradient at 10 months of age (f).