Use of a modified microvascular plug to modify fenestration flow in a rapidly failing Fontan

Published online by Cambridge University Press:  15 January 2024

Nicola McCay*
Affiliation:
Children’s Health Ireland, Dublin, Ireland
Philip Roberts
Affiliation:
Children’s Health Ireland, Dublin, Ireland
Kevin Walsh
Affiliation:
Children’s Health Ireland, Dublin, Ireland
*
Corresponding author: Nicola McCay; Email: Nicmccay29@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

We report a case of an alternative transcatheter use of the modified Medtronic microvascular plug to modify fenestration stent flow in a patient with a rapidly deteriorating clinical condition. This four-year-old boy developed severe cyanosis following fenestration stent insertion, initially placed due to prolonged drainage post-Fontan with extra-cardiac conduit. In April 2023, he underwent urgent cardiac catheterisation and had partial occlusion of fenestration stent with a modified 9Q microvascular plug. His oxygen saturations improved from 50 to 89% in room air with no re-emergence of raised cavopulmonary pressures.

Information

Type
Brief Report
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Figure 1. 1. Use of an ophthalmic cautery tool to create a hole in the PTFE covering of the microvascular plug. 2. The hole was balloon dilated with a 4 × 12 mm Maverick balloon. 3. A modified 9QMVP was made by the creation of a 4 mm hole to create a flow restrictor device.

Figure 1

Figure 2. m9QMVP in Cook F535 6 × 12 mm fenestration stent seen in AP and lateral angiographic images taken from the extra-cardiac conduit.