Efficiency and usability of an elementary hand-held suction assembly in paediatric tracheostomised patients

Published online by Cambridge University Press:  13 July 2022

P Sagar*
Affiliation:
Department of Otolaryngology and Head–Neck Surgery, All India Institute of Medical Sciences, New Delhi, India
R Kumar
Affiliation:
Department of Otolaryngology and Head–Neck Surgery, All India Institute of Medical Sciences, New Delhi, India
A Chandran
Affiliation:
Department of Otolaryngology and Head–Neck Surgery, All India Institute of Medical Sciences, New Delhi, India
*
Author for correspondence: Dr Prem Sagar, Department of Otolaryngology and Head–Neck Surgery, All India Institute of Medical Sciences, Ansari Nagar, New Delhi, Delhi 110029, India E-mail: sagardrprem@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Objective

Commercially available suction devices are expensive, large and heavy, and need electricity, and thus restrict the outdoor activity of tracheostomised children and their carers. This study evaluated the efficacy and usability of a simple suction assembly using a syringe and feeding tube in paediatric tracheostomised patients.

Methods

Following the domiciliary usage of this suction assembly instead of their existing suction device for a minimum of 15 days, carers responded to a set of questionnaires containing a subjective scoring system.

Results

Ninety-three per cent of the carers considered this assembly as average, good or very good in cleaning the tracheostomy tube. Eighty per cent of the carers considered that this assembly would be suitable when their existing suction machines are unavailable, indicating high usability, and 66.67 per cent of the carers would be confident using this assembly in outdoor settings.

Conclusion

Larger studies with objective evaluation methods can validate the high efficacy of this simple, inexpensive and easy-to-use, hand-held suction apparatus as reported by the carers of 15 paediatric tracheostomised patients in this study.

Information

Type
Main Article
Copyright
Copyright © The Author(s), 2022. Published by Cambridge University Press on behalf of J.L.O. (1984) LIMITED
Figure 0

Fig. 1. Hand-held suction assembly, consisting of a 50 ml syringe connected to a feeding tube of appropriate size and length.

Figure 1

Table 1. Predetermined feeding tube size and length, based on tracheostomy tube size*

Figure 2

Table 2. Parameters indicating efficiency and usability of suction assembly as rated by carers