Where in the cell is my protein?

Published online by Cambridge University Press:  21 June 2021

David J. DeRosier*
Affiliation:
Brandeis University, Waltham, MA, USA
*
Author for correspondence: David J. DeRosier, E-mail: derosier@brandeis.edu
Rights & Permissions [Opens in a new window]

Abstract

The application of cryo-correlative light and cryo-electron microscopy (cryo-CLEM) gives us a way to locate structures of interest in the electron microscope. In brief, the structures of interest are fluorescently tagged, and images from the cryo-fluorescent microscope (cryo-FM) maps are superimposed on those from the cryo-electron microscope (cryo-EM). By enhancing cryo-FM to include single-molecule localization microscopy (SMLM), we can achieve much better localization. The introduction of cryo-SMLM increased the yield of photons from fluorophores, which can benefit localization efforts. Dahlberg and Moerner (2021, Annual Review of Physical Chemistry, 72, 253–278) have a recent broad and elegant review of super-resolution cryo-CLEM. This paper focuses on cryo(F)PALM/STORM for the cryo-electron tomography community. I explore the current challenges to increase the accuracy of localization by SMLM and the mapping of those positions onto cryo-EM images and maps. There is much to consider: we need to know if the excitation of fluorophores damages the structures we seek to visualize. We need to determine if higher numerical aperture (NA) objectives, which add complexity to image analysis but increase resolution and the efficiency of photon collection, are better than lower NA objectives, which pose fewer problems. We need to figure out the best way to determine the axial position of fluorophores. We need to have better ways of aligning maps determined by FM with those determined by EM. We need to improve the instrumentation to be easier to use, more accurate, and ice-contamination free. The bottom line is that we have more work to do.

Information

Type
Review
Copyright
Copyright © The Author(s), 2021. Published by Cambridge University Press
Figure 0

Fig. 1. (a) Schematic diagram showing the insertion of a cylindrical lens in the optical path in an inverted microscope. The cylindrical lens introduces astigmatism into the PSF as seen in the through focus series of images to the right of the schematic drawing. At −400 nm, the PSF is elongated vertically in the image while at +400, it is elongated horizontally. (b) Plots of the horizontal width and vertical width versus distance from focus. The wx and wy widths of fluorophores in an image can be used to determine their z positions relative to the plane of focus. Reprinted from Fig. 1 by permission from © American Association for the Advancement of Science, Science, Huang et al. (2008).

Figure 1

Fig. 2. Schematic of a cryo-fluorescence microscope that can be used for SMLM. The 405 nm laser is typically used to switch fluorophores such as PSmOrange from orange to red. The 643 nm laser excites the switched PSmOrange fluorophore, which emits at 662 nm. The dichroic mirror to the left of the lasers transmits the excitation beam and reflects the activation beam making the two beams collinear. The other dichroic mirror reflects the laser beams onto the sample via the objective lens while transmitting the fluorescence beam returning from the sample. The returning beam is further isolated from the activation and excitation beams by the emission filter. From there the fluorescence beam is made stigmatic by a cylindrical lens, focused by the tube lens, and recorded by the camera. Additional lasers can be added using suitably chosen dichroic mirrors to merge the beams. To reflect these additional beams onto the sample and permit passage of the emitted fluorescent beam, we need to switch the second dichroic mirror and the emission filter to work with the altered wavelengths. The frozen-hydrated sample is kept frozen either by a cryo-block or by cold gas or liquid around the sample as indicated in one of the cryo-stage designs in Fig. 3.

Figure 2

Fig. 3. Schematics of four cold stage designs. (a) Cold nitrogen gas flows past the cryo-block, which holds the sample. A coverslip separates the cold sample from the objective lens. The objectives are limited to NAs <1. (b) In this design, a grid has been plunged and held in place by tweezers (not shown). The optical track runs horizontally from left to right passing through a window, the liquid propane, the sample, and then to the objective lens and camera (not shown). The liquid propane is maintained at the proper temperature by liquid nitrogen (not shown). The objective can have an NA of 1.3. (c) Cold nitrogen gas maintains vitreous nature of the ice in the sample. To permit the use of higher NAs, e.g. NA = 1.2, the sample and objective are optically coupled by a cryo-fluid, coverslip, and water. The cold stage operates at steady state. Heat, supplied to the water layer by some mechanism (not shown), is conducted through the coverslip and cryo-liquid to the sample. The heat is carried away by cold-nitrogen-gas cooled cryo-block on which the sample sits. The heat conductances of the coupling layers are such that the specimen is kept below the transition temperature of vitreous ice. (d) Cold nitrogen gas keeps the sample below the vitreous ice transition temperature. The sample sits under a high NA superSIL lens, which serves as the objective. Below the sample lies the virtual image of the sample made by the superSIL lens. The virtual image is reimaged by a second, longer-working- distance objective, which operates at ambient temperature. SuperSIL lenses with NAs above 2 are possible with this design.

Figure 3

Fig. 4. The top row of images of the 2D PSF is actual micrographs of fluorophores with dipoles that are fixed in space. The second row shows the simulated images of the 2D PSF using the measured polar and azimuthal angles of the dipole as determined for the top row of images. The azimuthal angles of the dipoles are indicated by the red arrows in the second row of images. The arrowhead on the arrow indicates the direction of the lobe of the dipole's power profile closest to the objective. The third row shows the polar tilt of the dipole (indicated here by the red arrow) from the plane of the specimen. The lobes flanking the arrow show a cross section of the distribution of radiated power from the dipole. In the images, the true position of the fluorophore is the center of the image, which is at x, y = 5, 5. With each pixel being 44 nm, it is easy to see that a Gaussian fit to images in (c), for example, would result in a significant shift of the calculated peak position from the true position of the fluorophore. Reprinted with modification from Fig. 1 by permission from © Nature/Springer/Palgrave, Nature Methods, Mortensen et al. (2010).

Figure 4

Fig. 5. The left side of the figure depicts an objective lens focused through a supporting film (horizontal blue line) onto a fluorophore (yellow burst) in the heart of a cell (dome-shaped structure) of refractive index n. The right side shows the objective focused on a fluorophore attached to the supporting film. The change in focus d multiplied by n gives the true position z of the fluorophore in the cell relative to the fluorophore on the grid.