Pulmonary valve-in-valve implantation using a Pulsta valve in repaired Tetralogy of Fallot

Published online by Cambridge University Press:  30 September 2025

Phuc Nang Vu*
Affiliation:
Cardiovascular Center, Tam Anh General Hospital, Ho Chi Minh City, Vietnam
Long Pham Hoang Nguyen
Affiliation:
Cardiovascular Center, Tam Anh General Hospital, Ho Chi Minh City, Vietnam
Gi Beom Kim
Affiliation:
Department of Pediatrics, Seoul National University Children’s Hospital, Seoul National University College of Medicine, Seoul, Korea
*
Corresponding author: Phuc Nang Vu; Email: nangphucvu@icloud.com
Rights & Permissions [Opens in a new window]

Abstract

We report the first case in Vietnam of transcatheter pulmonary valve-in-valve implantation using the self-expanding Pulsta valve in a young adult with repaired Tetralogy of Fallot. This case demonstrates the feasibility, safety, and early effectiveness of self-expanding valves in complex right ventricular outflow tract anatomies and highlights the importance of multimodal imaging in procedural planning.

Information

Type
Case Report
Copyright
© The Author(s), 2025. Published by Cambridge University Press
Figure 0

Figure 1. (a) Chest radiography showing cardiomegaly, dilated pulmonary artery trunk, left diaphragmatic paralysis, and mild scoliosis. (b) Electrocardiography showing sinus rhythm, complete right bundle branch block (QRS 162 ms), and right ventricular hypertrophy. (c–e) Transthoracic echocardiography showing the 23-mm trifecta valve, dilated and dysfunctional right ventricle, and hourglass-shaped right ventricular outflow tract. (f) Cardiac MRI assessing RVOT size for device selection.

Figure 1

Figure 2. (a) Pulmonary angiography showing hourglass-shaped RVOT with the trifecta bioprosthetic valve at the waist. (b) High-pressure balloon dilatation of the pulmonary valve and low-pressure balloon sizing to assess adequate expansion. (c, d) Deployment of a 24-mm pulsta valve (TaeWoong Medical, Gimpo, Republic of Korea) using an 18-fr delivery system. (e, f) Post-implantation assessment showing good valve expansion and angiographic confirmation of no intra- or para-valvular leak.