Procedural and short-term outcomes of transcatheter closure of ventricular septal defect using lifetech multifunctional occluder: initial experience

Published online by Cambridge University Press:  09 December 2020

Judah D. Gozar*
Affiliation:
Philippine General Hospital, Pediatrics, Taft Avenue, Metro Manila 1000, Philippines
Dexter E. D. Cheng
Affiliation:
Philippine General Hospital, Pediatrics, Taft Avenue, Metro Manila 1000, Philippines
Jose J. D. Del Rosario
Affiliation:
Philippine General Hospital, Pediatrics, Taft Avenue, Metro Manila 1000, Philippines
*
Author for correspondence: Judah Gozar, Philippine General Hospital, Pediatrics, Taft Avenue, Metro Manila 1000, Philippines. Tel: +63-9173100955. E-mail: drjudahtinagozar@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

The Lifetech Multifunctional occluder is a versatile device with an improved delivery and flexibility that reduces the risk of atrioventricular block. This is a retrospective, descriptive, pilot study done in 25 patients who underwent transcatheter closure of ventricular septal defect using Lifetech Multifunctional occluder from February 2017 to January 2018.

The average age was 9.32 ± 7.20 years, with a range from 1 to 32 years. Procedural success was 100% with no case needing a change of device size. Closure rate on follow up was at 42% (10/24), 52% (12/23), and 81% (17/21) at 1 day, 1 month, and 6 months, respectively. At 6-month follow up, pre-procedure tricuspid regurgitation disappeared by 38%. However, the incidence of new onset tricuspid regurgitation to trace was 16% (2) and mild 8% (1). Pre-procedure mild aortic regurgitation remained the same status throughout the 6-month follow up. Closure of the defect did not improve or worsen the aortic regurgitation. Post-transcatheter closure of ventricular septal defect with mild infundibular hypertrophy, the 1-year-old patient had resolution of the infundibular hypertrophy after 6 months but our 9-year-old patient had persistence of the mild infundibular hypertrophy even after 6 months. One patient (4%) developed transient widened QRS complexes post-transcatheter closure that resolved after 1 month. In total, 92% of the patients had no periprocedural complications. While one patient each had an inadvertent urinary bladder puncture and device embolisation.

Our retrospective review of the procedural and short-term outcomes of transcatheter closure of ventricular septal defect sizes 2–10 mm, using the Lifetech Multifunctional occluder, appears to be safe and effective. However, long-term follow up is warranted.

Information

Type
Original Article
Copyright
© The Author(s), 2020. Published by Cambridge University Press
Figure 0

Figure 1. The Lifetech MFO VSD Occluder.

Figure 1

Table 1. Specification of Lifetech Multifunctional occluder (MFO)

Figure 2

Table 2. Baseline and procedural characteristics of patients with VSD who underwent transcatheter closure using Lifetech MFO

Figure 3

Figure 2. Steps in Antegrade approach. (a) Left ventricular angiogram showing the VSD. (b) The VSD is crossed retrogradely from the left ventricle using a modified cut pigtail or judkins right catheter and a 260 cm glide wire, which is advanced into the main pulmonary artery or superior vena cava. The wire is snared from a femoral vein approach and externalized creating an arteriovenous wire loop. (c) The delivery sheath is advanced from the femoral vein through the VSD and into the descending aorta. The wire is removed. (d) The defect is measured using angiographic measurements. The device was chosen to be exactly the same size to up to 2 mm larger than the right ventricular side defect as measured on transthoracic echocardiogram and/or angiogram. The device was advanced through the sheath and deployed across the VSD in a standard fashion. (e) Repeat left ventricular angiogram confirms the placement of the device within the defect. Aortogram is done to assess aortic regurgitation and transthoracic echocardiography to assess tricuspid regurgitation. (f) The device is released from the delivery system.

Figure 4

Figure 3. Steps in Retrograde approach. (a) Left ventricular angiogram showing the VSD. The defect is measured using angiographic measurements and the device was chosen to be exactly the same size to up to 2 mm larger than the right ventricular side defect as measured on transthoracic echocardiogram and/or angiogram. (b) The VSD is crossed retrogradely from the left ventricle using a modified cut pigtail or judkins right catheter and a 260 cm glide wire, which is advanced into the right ventricle. The pigtail or judkins right catheter is removed and the delivery sheath is advanced retrogradely from the femoral artery through the VSD and into the right ventricle. The wire is removed. (c) The device is advanced through the sheath and deployed across the VSD in a retrograde fashion. Transthoracic echocardiography is done to assess tricuspid regurgitation. The device is released from the delivery system.

Figure 5

Table 3. Complete closure rate of the defect at 1 day, 1 month, and 6 months post-transcatheter closure

Figure 6

Figure 4. TTE 5 chamber view showing a left to right shunting through the perimembranous VSD.

Figure 7

Figure 5. TTE 5 chamber view showing a Lifetech MFO with complete occlusion of the defect.

Figure 8

Table 4. Incidence of pre-procedure tricuspid regurgitation

Figure 9

Table 5. Course of pre-procedure tricuspid regurgitation

Figure 10

Table 6. Incidence of new tricuspid regurgitation

Figure 11

Figure 6. Post transcatheter closure tracing showing widened QRS complexes in all limb leads.

Figure 12

Figure 7. TTE short axis view showing a perimembranous VSD with mild infundibular pulmonary stenosis with a gradient of 38 mmHg.

Figure 13

Figure 8. TTE subcostal view showing a perimembranous VSD with mild infundibular pulmonary stenosis with a gradient of 38 mmHg.

Figure 14

Figure 9. Left ventriculogram in LAO Cranial view showing an egress of dye through the perimembranous VSD to the right ventricle with mild infundibular pulmonary stenosis.

Figure 15

Figure 10. TTE short axis view showing a perimembranous VSD with mild infundibular pulmonary stenosis with a gradient of 26 mmHg.

Figure 16

Figure 11. Angiogram in the right ventricular outflow tract in lateral view showing a mild infundibular stenosis.

Figure 17

Figure 12. Gerbode Defects. The type A or Direct type, the blood shunts from the left ventricle through the membranous septum above the tricuspid valve into the right atrium. The type B Gerbode, or Indirect type, the blood flow from the left ventricle shunting through the VSD into the right ventricle and then through the septal leaflet defect of the tricuspid valve into the right atrium. Image taken from Vijayalakshmi IB. Device Closure of Small Ventricular Septal Defects: When and Why? BMH Medical Journal (ISSN 2348-392X), 1(3): 56-63 (2014).

Figure 18

Figure 13. TTE 4 chamber view showing a perimembranous VSD with shunting of the flow through the septal tricuspid leaflet causing tricuspid regurgitation.

Figure 19

Figure 14. TTE 4 chamber view showing a severe tricuspid regurgitation from a type B or Indirect Gerbode defect.

Figure 20

Figure 15. TTE 5 chamber view showing occlusion of the Geborde type B defect with no residual shunt.

Figure 21

Figure 16. TTE 4 chamber view showing resolution of the tricuspid regurgitation after occlusion of the Gerbode defect.