A tangled affair: pacemaker malfunction and syncope in a child due to Twiddler's syndrome

Published online by Cambridge University Press:  23 January 2007

Alexander R. Ellis
Affiliation:
Division of Pediatric Cardiology, Children's Hospital, Medical University of South Carolina, Charleston, South Carolina, United States of America
John H. Reed
Affiliation:
Division of Pediatric Cardiology, Children's Hospital, Medical University of South Carolina, Charleston, South Carolina, United States of America
David Fairbrother
Affiliation:
Division of Pediatric Cardiology, Children's Hospital, Medical University of South Carolina, Charleston, South Carolina, United States of America
Rights & Permissions [Opens in a new window]

Abstract

Manipulation of an implanted pacemaker by the patient is a rare cause of malfunction, especially in children. We describe a child who inadvertently rotated his pacemaker under the skin, knotting the leads and dislodging them from the heart, leading to syncope and heart block. Our experience with this case underscores the need to consider this diagnosis in children as well as in adults if this problem is to be averted.

Information

Type
Brief Report
Copyright
© 2007 Cambridge University Press
Figure 0

(a) Baseline chest X-ray, showing normally-placed epicardial leads and the device positioned within the chest. (b) X-ray taken after the malfunction, showing knotted epicardial leads and inversion of the device on its long-axis. (c) Re-implantation of a new generator in an abdominal position, along with new screw-in leads and explantation of the previous displaced generator and knotted leads.

Figure 1

A 21 second episode ventricular non-capture. Episodes of non-capture could be precipitated by turning this patient left side up, which removed the pacemaker leads from the connection with the epicardial surface.