Correlation of thickness changes and daily dose detection of delivery analysis (DA) software in helical tomotherapy

Published online by Cambridge University Press:  17 October 2024

Nattawut Wijarn
Affiliation:
Medical Physics Program, Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
Anirut Watcharawipha
Affiliation:
The Division of Radiation Oncology, Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
Wannapha Nobnop*
Affiliation:
The Division of Radiation Oncology, Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand
*
Corresponding author: Wannapha Nobnop; Email: pung435@yahoo.com
Rights & Permissions [Opens in a new window]

Abstract

Purpose:

To assess the correlation between the dose distribution provided by the delivery analysis (DA) software and the measured dose distribution using an ArcCHECK (AC) phantom in the presence of thickness variation.

Materials and Methods:

Two sets of targets were established within the phantom. Target A was placed on the detector areas, whereas Target B was positioned at the centre of the phantom. Bolus was applied to the surface of the phantom at different thicknesses ranging from 0 to 2 cm to verify the dose distribution in both TomoHelical (HT) and TomoDirect (TD) techniques. The gamma passing rate (GPRs) were evaluated against predefined thresholds of 3%/3 mm and 3%/2 mm. Correlation study evaluated the level of agreement between DA and AC values.

Results:

Both AC and DA exhibited a decline in GPRs as the bolus thickness decreased. Significant correlations were observed between AC and DA for both HT and TD techniques, with a p value of less than 0.001.

Conclusion:

The results indicate that DA software has the capability to detect anatomical changes during tomotherapy treatments. This is substantiated by the statistical association found between DA and the standard AC phantom system for dose distributions in both HT and TD methods.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BY
This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution and reproduction, provided the original article is properly cited.
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Figure 1. Diagram of the study design.

Figure 1

Figure 2. Dose distribution on ArcCHECK phantom in TomoHelical planning for the Target A plans (a) on the target volume (a1) and organ at risk: OAR volume (a2) and the Target B plans (b) on the target volume (b1) and organ at risk: OAR volume (b2).

Figure 2

Table 1. The correlation of gamma passing rates between ArcCHECK phantom (AC) and delivery analysis (DA) software with thickness variation (0–2 cm) for TomoHelical plans

Figure 3

Figure 3. Dose distribution on ArcCHECK phantom in TomoDirect planning in 0, 40 and 320 degrees for the Target A plans (a) on the target volume (a1) and organ at risk: OAR volume (a2) and the Target B plans (b) on the target volume (b1) and organ at risk: OAR volume (b2).

Figure 4

Table 2. The correlation of gamma passing rates between ArcCHECK phantom (AC) and delivery analysis (DA) software with thickness variation (0–2 cm) for TomoDirect plans