Patent ductus arteriosus closure in small infants weighing less than 1500 g using KA micro plug: steps and tips

Published online by Cambridge University Press:  04 October 2024

Bassel Mohammad Nijres*
Affiliation:
Pediatric Cardiology Section, Department of Pediatrics, King Faisal Specialist Hospital and Research Center, Jeddah, Saudi Arabia
Moeez Hassan
Affiliation:
Division of Pediatric Cardiology, Stead Family Children’s Hospital, University of Iowa, Iowa City, IA, USA
Jimmy Windsor
Affiliation:
Division of Pediatric Anesthesia, Stead Family Children’s Hospital, University of Iowa, Iowa City, IA, USA
Adrianne Rahde Bischoff
Affiliation:
Division of Neonatology, Stead Family Children’s Hospital, University of Iowa, Iowa City, IA, USA
Patrick J. McNamara
Affiliation:
Division of Neonatology, Stead Family Children’s Hospital, University of Iowa, Iowa City, IA, USA
Osamah Aldoss
Affiliation:
Division of Pediatric Cardiology, Stead Family Children’s Hospital, University of Iowa, Iowa City, IA, USA
*
Corresponding author: Bassel Mohammad Nijres; Email: bnijres@kfshrc.edu.sa
Rights & Permissions [Opens in a new window]

Abstract

Background:

Transcatheter patent ductus arteriosus closure in premature infants is gaining attention with rapidly growing experience. The KA micro plug device poses many favourable features for ductus arteriosus closure in premature infants. The study aimed to describe the technique and outcomes of a KA micro plug in closing the ductus arteriosus in premature infants weighing less than 1500 g.

Methods:

This is a single-centre retrospective study in premature infants with patent ductus arteriosus weighing less than 1500 g who underwent off-label use of a KA micro plug at the Stead Family Children’s Hospital from February 2022 to December 2023.

Results:

Eighteen premature infants weighing less than 1500 g underwent attempted transcatheter ductus arteriosus closure using a KA micro plug device. Median weight, age, and corrected gestational age at the procedure were 943 g (682–1225), 26 days (9–79), and 28.5 weeks (25.6–32), respectively. The procedure was successful with complete closure achieved in all patients. No major haemodynamic instabilities were encountered. No catheterization-related complications were encountered. On a median follow-up of 10.9 months (0.1–19), all patients are alive with the device in a good position without residual shunt or abnormal findings.

Conclusion:

KA micro plug device for patent ductus arteriosus in infants weighing less than 1500 g is feasible, safe, and effective. We recommend that manufacturers of devices used to close the ductus arteriosus in small infants enhance both the devices and their delivery systems to simplify the procedure and mitigate the risk of haemodynamic instability.

Information

Type
Original Article
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Figure 1. Images demonstrate the steps of using the KA micro plug. A. Marking the Angled Glide catheter length on the drape of the back table. B. Cutting the microcatheter 12 cm longer than the Angled Glide catheter. C. Front loading and de-airing the KA device inside a heparinised saline bowel. D. Pulling the Angled Glide catheter (single white arrow) over the microcatheter (double white arrows) to expose the distal end of the microcatheter. Notice that the Y-connectors (black arrows) are connected with the Angled Glide catheter and the microcatheter. E. Fixing the Angled Glide catheter (single white arrow) and the microcatheter (double white arrows) with the left hand and fixing the delivery wire (black arrow) with the left hand. The device is deployed by pulling both catheters as one unit over the device wire. F. Showing the KA micro plug device is deployed. Notice that when the Angled Glide catheter (single white arrow) is pulled over the microcatheter (double white arrow) as shown in “D,”, the microcatheter extends approximately 2 cm outside the distal end of the Angled Glide catheter.

Figure 1

Figure 2. Images demonstrate the utility of intraprocedural angiogram and echocardiogram. A. Patent ductus arteriosus angiogram in the lateral projection with caudal angulation depicts measuring patent ductus arteriosus’s dimensions. B and C. Main pulmonary artery angiogram after KA micro plug deployment shows the device inside the patent ductus arteriosus (white arrow) with no left pulmonary artery obstruction. B. Lateral projection with caudal angulation. C. Frontal projection with left anterior oblique angulation. D. Still echocardiogram image shows the KA micro plug (arrow) inside the patent ductus arteriosus without residual shunt.

Figure 2

Table 1. Summarising demographics

Figure 3

Table 2. Summarising patent ductus arteriosus dimensions and types

Supplementary material: File

Mohammad Nijres et al. supplementary material 1

Mohammad Nijres et al. supplementary material
Download Mohammad Nijres et al. supplementary material 1(File)
File 10.4 MB
Supplementary material: File

Mohammad Nijres et al. supplementary material 2

Mohammad Nijres et al. supplementary material
Download Mohammad Nijres et al. supplementary material 2(File)
File 1.1 MB