Complete atrioventricular block after atrial flow regulator implantation in a 5 kg infant

Published online by Cambridge University Press:  03 October 2024

Céline Grunenwald Gronier
Affiliation:
Nantes Université, CHU Nantes, Department of Pediatric Cardiology and Pediatric Cardiac Surgery, FHU PreciCare, Nantes, France
Nadir Benbrik
Affiliation:
Nantes Université, CHU Nantes, Department of Pediatric Cardiology and Pediatric Cardiac Surgery, FHU PreciCare, Nantes, France
Alban-Elouen Baruteau*
Affiliation:
Nantes Université, CHU Nantes, Department of Pediatric Cardiology and Pediatric Cardiac Surgery, FHU PreciCare, Nantes, France Nantes Université, CHU Nantes, INSERM, CIC FEA 1413, Nantes, France Nantes Université, CHU Nantes, CNRS, INSERM, l’institut du thorax, Nantes, France Nantes Université, INRAE, UMR 1280, PhAN, Nantes, France
*
Corresponding author: Alban-Elouen Baruteau; Email: albanelouen.baruteau@chu-nantes.fr
Rights & Permissions [Opens in a new window]

Abstract

The atrial flow regulator is a new self-expandable double-disc fenestrated device providing a calibrated inter-atrial communication. Paediatric reports are scarce. We herein describe a case of complete atrioventricular block complicating the exemption use of an atrial flow regulator in a 5-kg infant with transposition of the great arteries, ventricular septal defect, and right ventricular outflow tract obstruction.

Information

Type
Brief Report
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Figure 1. Echocardiography-guided transseptal puncture with a Brockenbrough needle (a). Contrast injections through the sheath confirm that the puncture site is located in the middle of the inter-atrial septum, 12 mm away from the mitral valve (b, c). Static balloon dilation using a 12x20 mm Powerflex balloon catheter (d) and a 14x20 mm Atlas Gold balloon catheter (e) on a 0.035’ guidewire placed in the left upper pulmonary vein. After percutaneous implantation of a 18/6 atrial flow regulator, caudal right anterior oblique view (i.e., enface view) shows a well-opened 6 mm fenestration (f).

Figure 1

Figure 2. Electrocardiographic monitoring showing occurrence of complete atrioventricular dissociation.

Supplementary material: File

Grunenwald Gronier et al. supplementary material 1

Grunenwald Gronier et al. supplementary material
Download Grunenwald Gronier et al. supplementary material 1(File)
File 884.8 KB
Supplementary material: File

Grunenwald Gronier et al. supplementary material 2

Grunenwald Gronier et al. supplementary material
Download Grunenwald Gronier et al. supplementary material 2(File)
File 630.6 KB
Supplementary material: File

Grunenwald Gronier et al. supplementary material 3

Grunenwald Gronier et al. supplementary material
Download Grunenwald Gronier et al. supplementary material 3(File)
File 12.6 KB