The impact of reference isocentre position on set-up errors in head-and-neck image-guided radiotherapy

Published online by Cambridge University Press:  02 October 2017

Helena Lenko
Affiliation:
Institute of Oncology Ljubljana, Ljubljana, Slovenia
Primož Peterlin*
Affiliation:
Institute of Oncology Ljubljana, Ljubljana, Slovenia
*
Correspondence to: Primož Peterlin, Institute of Oncology Ljubljana, Zaloška c. 2, SI-1000 Ljubljana, Slovenia. Tel: +386 1 5879509. E-mail: ppeterlin@onko-i.si
Rights & Permissions [Opens in a new window]

Abstract

Aim

To examine and quantify set-up errors in patient positioning in head-and-neck radiotherapy and to investigate the impact of the choice of reference isocentre—on the patient neck or patient skull—on the magnitude of set-up errors.

Materials and methods

Set-up position corrections obtained using online kV 2D/2D matching were recorded automatically for every treatment fraction. 3,413 treatment records for 117 patients treated with volumetric modulated arc therapy during 2013 and 2014 on a single treatment machine in our clinic were analysed. In 79 treatment plans the reference isocentre was set to the patient skull, and in 47 to the neck.

Results

Standard deviation of group systematic error in the vertical, longitudinal and lateral direction and the couch rotation were found to be 2·5 mm, 2·1 mm, 1·9 mm and 0·43° (skull) and 2·5 mm, 1·8 mm, 1·7 mm and 0·49° (neck), respectively. Random error of the vertical, longitudinal, lateral and rotational position correction was 1·8 mm, 1·5 mm, 1·6 mm and 0·62° (skull) and 1·9 mm, 1·6 mm, 1·5 mm and 0·60° (neck), respectively. Positional shifts in different directions were found to be uncorrelated.

Conclusions

Neither reference isocentre set-up shows a clear advantage over the other in terms of interfraction set-up error.

Information

Type
Original Articles
Copyright
© Cambridge University Press 2017 
Figure 0

Figure 1 Distribution of longitudinal shifts from the reference isocentre to treatment field isocentre for patients with reference isocentre set on patient skull (red) and neck (blue). Positive and negative values denote shifts in the cranial and caudal directions, respectively.

Figure 1

Table 1 Systematic and random errors in patient position with respect to the position of reference isocentre

Figure 2

Figure 2 Histograms of distributions of corrections applied in the vertical (a), longitudinal (b) and lateral (c) direction, and couch rotation corrections (d). The red lines denote the patients with reference isocentre on patient skull, blue the patients with patient origin on patient neck.

Figure 3

Figure 3 Histogram of the distribution of the magnitude of translational positional corrections applied (a); the correlation between the magnitude of translational correction and rotational correction (b). Red denotes the patients with reference isocentre on the skull, while blue the patients with reference isocentre on the neck.

Figure 4

Figure 4 Correlations between positional corrections in different directions. An individual circle in a scatterplot corresponds to the positional corrections performed in different directions at an individual treatment fraction. Red and blue circles (blue plotted on top of red) denote the patients with reference isocentre on patient skull and patient neck, respectively. The absolute value of Kendall rank correlation coefficient (τB) and the corresponding p-value for the patients with reference isocentre on the skull and on the neck are written in red and blue, respectively.

Figure 5

Table 2 Comparison of the values of systematic (Σ) and random error (σ) in the vertical, longitudinal and lateral direction, and couch rotation, in a few recent studies analysing set-up errors in treating head-and-neck cancer