Hostname: page-component-74656f9884-rdgjk Total loading time: 0 Render date: 2026-11-07T09:13:28.590Z Has data issue: false hasContentIssue false

Spatially-controlled illumination microscopy

For prolonged live-cell and live-tissue imaging with extended dynamic range

Published online by Cambridge University Press:  12 December 2016

Venkataraman Krishnaswami
Affiliation:
Core Facility Cellular Imaging, van Leeuwenhoek Centre for Advanced Microscopy (LCAM), Academic Medical Centre (AMC), University of Amsterdam, Meibergdreef 15, 1105 AZ Amsterdam, The Netherlands van Leeuwenhoek Centre for Advanced Microscopy (LCAM), Swammerdam Institute for Life Sciences (SILS), University of Amsterdam, Science Park 904, 1098 XH Amsterdam, The Netherlands
Cornelis J. F. Van Noorden
Affiliation:
Core Facility Cellular Imaging, van Leeuwenhoek Centre for Advanced Microscopy (LCAM), Academic Medical Centre (AMC), University of Amsterdam, Meibergdreef 15, 1105 AZ Amsterdam, The Netherlands
Erik M. M. Manders
Affiliation:
van Leeuwenhoek Centre for Advanced Microscopy (LCAM), Swammerdam Institute for Life Sciences (SILS), University of Amsterdam, Science Park 904, 1098 XH Amsterdam, The Netherlands
Ron A. Hoebe*
Affiliation:
Core Facility Cellular Imaging, van Leeuwenhoek Centre for Advanced Microscopy (LCAM), Academic Medical Centre (AMC), University of Amsterdam, Meibergdreef 15, 1105 AZ Amsterdam, The Netherlands
*
*Author for correspondence: Ron A. Hoebe, Core Facility Cellular Imaging, van Leeuwenhoek Centre for Advanced Microscopy (LCAM), Academic Medical Centre (AMC), University of Amsterdam, Meibergdreef 15, 1105 AZ Amsterdam, The Netherlands. Tel.: 3156664743; Email: r.a.hoebe@amc.uva.nl
Rights & Permissions [Opens in a new window]

Abstract

Live-cell and live-tissue imaging using fluorescence optical microscopes presents an inherent trade-off between image quality and photodamage. Spatially-controlled illumination microscopy (SCIM) aims to strike the right balance between obtaining good image quality and minimizing the risk of photodamage. In traditional imaging, illumination is performed with a spatially-uniform light dose resulting in spatially-variable detected signals. SCIM adopts an alternative imaging approach where illumination is performed with a spatially-variable light dose resulting in spatially-uniform detected signals. The actual image information of the biological specimen in SCIM is predominantly encoded in the illumination profile. SCIM uses real-time spatial control of illumination in the imaging of fluorescent biological specimens. This alternative imaging paradigm reduces the overall illumination light dose during imaging, which facilitates prolonged imaging of live biological specimens by minimizing photodamage without compromising image quality. Additionally, the dynamic range of a SCIM image is no longer limited by the dynamic range of the detector (or camera), since it employs a uniform detection strategy. The large dynamic range of SCIM is predominantly determined by the illumination profile, and is advantageous for imaging both live and fixed biological specimens. In the present review, the concept and working mechanisms of SCIM are discussed, together with its application in various types of optical microscopes.

Information

Type
Review
Copyright
Copyright © Cambridge University Press 2016 
Figure 0

Fig. 1. Concept of SCIM in comparison with conventional optical microscopy. (a) Conventional optical microscopy uses spatially-uniform illumination to acquire a variable detection. (b) SCIM aims at spatially-uniform detection profiles in the background and foreground regions in accordance with the desired SNR by adopting spatial control of illumination light dose. It ‘turns ON’ illumination with sufficient light dose in the foreground regions and ‘turns OFF’ or minimizes illumination in the background regions. The actual image containing spatial intensity information of the specimen is reconstructed with mathematical estimation techniques using the detection and illumination profiles.

Figure 1

Fig. 2. Closed-loop feedback system for spatial control of illumination in a SCIM setup.

Figure 2

Fig. 3. Illustrative scheme of SCIM implementation in a CLSM. CLSM is a point-scanning imaging technique which employs a pinhole to enable optical sectioning. SCIM incorporates real-time monitoring and decision making at the pixel level, during acquisition with the help of feedback electronics. The feedback electronics then controls the AOM shutter to ‘turn ON’ and ‘turn OFF’ illumination to achieve spatial control of illumination light dose in background and foreground regions.

Figure 3

Fig. 4. Time-lapse images comparing conventional imaging with SCIM implemented in a CLSM introduced as CLEM as reproduced from Hoebe et al. (2007). Scale bar in (a) & (b), 5 µm. Scale bar in (c) & (d), 10 µm. (a) Fluorescence imaging of 3D stacks of tobacco BY-2 cells expressing microtubule-associated GFP-MAP4 with spatially-uniform illumination results in photobleaching within 25 min. (b) Fluorescence imaging using spatially-controlled illumination, of tobacco BY-2 cells results in much lesser photobleaching and prolongs imaging. (c) Conventional time-lapse imaging of living H2B-GFP-expressing HeLa cells in which the GFP signal (indicated in green) is superimposed on the phase-contrast images (indicated in grey). Yellow arrows indicate blebbing of the membrane after 27 min of exposure. After 96 min, the rounding up of cells begins as indicated by the orange arrows and after 169 min some cells are apoptotic as indicated by the red arrow. (d) Phase-contrast imaging of live HeLa cells after using SCIM in CLSM shows reduced photodamage, which allows prolonged live-cell imaging by a factor 6.

Figure 4

Fig. 5. Imaging of bovine endothelial cells showing actin labeled with Texas Red-X phalloidin using SCIM implemented in AIM in a two-photon microscope as reproduced from Chu et al. (2007). Scale bar, 20 µm. (a) Image recorded with the AIM two-photon setup has a constant SNR in background and foreground regions. (b) Illumination profile is spatially-controlled during imaging, where illumination is minimized in strong fluorescent regions of the specimen and enhanced in weak fluorescent regions. (c) Mathematically reconstructed image using the detected and the illumination profiles. (d) Image obtained by conventional two-photon imaging. The mathematically reconstructed image in (c) is almost identical to the image obtained by conventional two-photon imaging in (d).

Figure 5

Fig. 6. Illustrative scheme of a wide-field dual path PAM. PAM uses a SLM such as a DMD array build in the setup of a dual path PAM for spatial control of illumination patterns that can be generated and programmed dynamically during imaging. A dual path PAM records both the conjugate and non-conjugate images obtained with the DMDs in ‘ON’ and ‘OFF’ positions, respectively. Optical sectioning is also achieved by mathematical treatment of recorded conjugate and non-conjugate images. The implementation of SCIM in PAM is called MLE-PAM and can be considered as a wide-field-based, discretized version of SCIM in CLSM, where decisions for all pixels are taken at the same time in each image at constant time intervals. The illumination is then controlled by dynamically programming the patterns of a DMD array in accordance to the illumination profile required in background and foreground regions.

Figure 6

Fig. 7. Reduced photodamage in live-cell imaging of HeLa cells stained with microtracker to visualize mitochondria with MLE-PAM as reproduced from Caarls et al. (2011). Scale bar 5 µm. (a) Conventional time-lapse imaging using wide-field PAM shows significant photodamage within 13·11 min. (b) Time-lapse imaging using SCIM and PAM (MLE-PAM) shows reduced photodamage, which allows prolonged time-lapse imaging beyond 13·11 min.

Figure 7

Fig. 8. Illustrative scheme of an adaptive SIM. Adaptive SIM uses a DMD array in its construction. DMD arrays are used to generate patterned (based on a binary mask), multispot illumination over one specimen. SCIM is performed with SIM by using an additional greyscale mask comprising of weighing factors by which the intensity of the illumination is scaled. Spatially controlled illumination patterns are generated by combining the two masks. Greyscale masks are updated after every single imaging sequence based on preset thresholds and illumination schemes. Series of images are obtained sequentially using different illumination patterns and are mathematically reconstructed to generate a SIM image.

Figure 8

Fig. 9. Reduced photodamage in live-cell imaging of HEK 293T cells, in which DNA is labelled with SYTOX using adaptive SIM as reproduced from Chakrova et al. (2016a). Comparison of photobleaching effects over imaging time using standard SIM and adaptive SIM with linear and non-linear masks is shown. Scale bar, 2 µm. (a) First acquired image. (b) Image acquired after 10 min from the 20th acquisition. (c) Image acquired after 20 min from the 40th acquisition. (d) Image acquired after 30 min from the 60th acquisition. Upon 30 min of imaging time, the specimen is subjected to heavy photobleaching effects when imaged using a standard SIM setup. In comparison, photobleaching effects are significantly lower when imaged using adaptive SIM, which incorporates the principle of SCIM in its operation.

Figure 9

Fig. 10. Illustrative scheme of RESCue STED microscopy. STED employs two illumination sources, which are combined to generate a doughnut-shaped illumination profile, which is used to ‘turn ON’ the fluorescent probes at the center with the excitation beam and ‘turn OFF’ all the adjacent fluorescent probes with the depletion beam to achieve super-resolution. SCIM in STED (RESCue STED) is implemented using a real-time feedback unit controlling the AOTF shutter which is used to ‘turn ON’ and ‘turn OFF’ the illumination beams when required.

Figure 10

Fig. 11. Imaging of fluorescently labelled proteins in fixed primary mouse neurons using STED nanoscopy and reduced numbers of excitation and signal suppression cycles (RESCue) STED as reproduced from Staudt et al. (2011). Scale bar, 1 µm. (a) Conventional imaging with spatially-uniform illumination using CLSM and STED nanoscopy with a 50 nm spatial resolution. (b) Conventional imaging with uniform illumination results in photodamage. (c) Imaging with foreground illumination strategy of SCIM using a fixed threshold of ‘ITh’, reduces photodamage. (d) Imaging with both the foreground and background strategy of SCIM using background threshold ‘uTh’ and foreground threshold ‘ITh’, provides more photodamage reduction. *(1) & (2) shows the images of a fluorescent sample captured with a confocal and a STED microscope respectively, before scanning. (3) & (4) shows the images after scanning, using confocal and STED microscopes, respectively.

Figure 11

Fig. 12. Prominent fluorescence-based optical microscopes that are available for live-cell and/or live-tissue imaging. SCIM has been implemented in both confocal and wide-field systems with optical sectioning capabilities.