Monte Carlo calculations of an Elekta Precise SL-25 photon beam model

Published online by Cambridge University Press:  18 August 2015

Fátima Padilla-Cabal
Affiliation:
Nuclear Physics Department, Instituto Superior de Tecnologías y Ciencias Aplicadas, La Habana, Cuba
Mailyn Pérez-Liva*
Affiliation:
Grupo de Física Nuclear, Facultad de Ciencias Físicas, Universidad Complutense de Madrid, Madrid, Spain
Elier Lara
Affiliation:
Department of Radiotherapy, Instituto Nacional de Oncología y Radiobiología, La Habana, Cuba
Rodolfo Alfonso
Affiliation:
Nuclear Physics Department, Instituto Superior de Tecnologías y Ciencias Aplicadas, La Habana, Cuba
Neivy Lopez-Pino
Affiliation:
Nuclear Physics Department, Instituto Superior de Tecnologías y Ciencias Aplicadas, La Habana, Cuba
*
Correspondence to: Mailyn Pérez-Liva, Ave Complutense, S/N, Grupo de Física Nuclear, Facultad de Ciencias Físicas, Universidad Complutense de Madrid, 28040 Madrid, Spain. Tel: +34 91 394 4484. Fax: +34 91 394 5193. E-mail: mailyn01@ucm.es
Rights & Permissions [Opens in a new window]

Abstract

Background

Monte Carlo (MC) simulations have been used extensively for benchmarking photon dose calculations in modern radiotherapy using linear accelerators (linacs). Moreover, a major barrier to widespread clinical implementation of MC dose calculation is the difficulty in characterising the radiation source using data reported from manufacturers.

Purpose

This work aims to develop a generalised full MC histogram source model of an Elekta Precise SL-25 linac (electron exit window, target, flattening filter, monitor chambers and collimators) for 6 MV photon beams used in standard therapies. The inclusion of many different probability processes such as scatter, nuclear reactions, decay, capture cross-sections and more led to more realistic dose calculations in treatment planning and quality assurance.

Materials and methods

Two different codes, MCNPX 2·6 and EGSr-BEAM, were used for the calculation of particle transport, first in the geometry of the internal/external accelerator source, and then followed by tracking the transport and energy deposition in phantom-equivalent tissues. A full phase space file was scored directly above the upper multilayer collimator’s jaws to derive the beam characteristics such as planar fluence, angular distribution and energy spectrum. To check the quality of the generated photon beam, its depth dose curves and cross-beam profiles were calculated and compared with measured data.

Results

In-field dose distributions calculated using the accelerator models were tuned to match measurement data with preliminary calculations performed using the accelerator information provided by the manufacturer. Field sizes of 3×3, 5×5, 10×10, 15×15 and 20×20 cm2 were analysed. Local differences between calculated and measured curve doses beneath 2% were obtained for all the studied field sizes. Higher discrepancies were obtained in the air–water interface, where measurements of dose distributions with the ionisation chamber need to be shifted for the effective point of measurement.

Conclusion

The agreements between MC-calculated and measured dose distributions were excellent for both codes, showing the strength and stability of the proposed model. Beam reconstruction methods as direct input to dose-calculation codes using the recorded histograms can be implemented for more accurate patient dose estimation.

Information

Type
Technical Note
Copyright
© Cambridge University Press 2015 
Figure 0

Figure 1 MCNPX view of the gamma linac head model.

Figure 1

Figure 2 MCNPX sketch of both phantoms used for beam dose characterisation.

Figure 2

Figure 3 Energy spectra of primary photons (a) and electrons (b) at the PS of a 6 MV beam from an Elekta Precise SL-25 accelerator (SSD=30 cm) using the MCNPX 2·6 code. Abbreviations: PS, phase space; SSD, source and the phantom surface.

Figure 3

Figure 4 Zenithal angular distribution of primary photons (a) and electrons (b) at the PS of a 6 MV beam from an Elekta Precise SL-25 accelerator (SSD=30 cm) using the MCNPX 2·6 code. Abbreviations: PS, phase space; SSD, source and the phantom surface.

Figure 4

Figure 5 Radial distribution of primary photons (a) and electrons (b) at the phase space of a 6 MV beam from an Elekta Precise SL-25 accelerator (SSD=30 cm) using the MCNPX 2·6 code. Abbreviation: SSD, source and the phantom surface.

Figure 5

Figure 6 Energy distribution of primary photons and electrons at the phantom surface of a 6 MV beam from an Elekta Precise SL-25 accelerator (SSD=100 cm) using the MCNPX 2·6 code. Abbreviation: SSD, source and the phantom surface.

Figure 6

Figure 7 Zenithal angular distribution of primary photons and contaminant electrons at the phantom surface of a 6 MV beam from an Elekta Precise SL-25 accelerator (SSD=100 cm) using the MCNPX 2·6 code. Abbreviations: FWHM, full-width at half-maximum; SSD, source and the phantom surface.

Figure 7

Figure 8 Radial distribution of primary photons and electrons at the phantom surface of a 6 MV beam from an Elekta Precise SL-25 accelerator (SSD=100 cm) using the MCNPX 2·6 code. Abbreviations: FWHM, full-width at half-maximum; SSD, source and the phantom surface.

Figure 8

Table 1 Average energy and scatter angle for primary and scattered radiation for a 6 MV Elekta Precise SL-25 photon beam determined on the beam axis calculated with the MCNPX 2·6 code

Figure 9

Figure 9 Radial distribution of primary photons (a) and electrons (b) at the phantom surface of a 6 MV beam from an Elekta Precise SL-25 accelerator (SSD=100 cm) using the MCNPX 2·6 code for three different square fields. Abbreviations: FWHM, full-width at half-maximum; SSD, source and the phantom surface.

Figure 10

Figure 10 Comparison between measured and calculated depth dose distribution at the central axis of a water phantom caused by a 6 MV beam from an Elekta Precise SL-25 accelerator (SSD=100 cm) using the MCNPX 2·6 and the EGS-BEAM codes. Abbreviations: FWHM, full-width at half-maximum; PDD, per cent dose depth; SSD, source and the phantom surface.

Figure 11

Figure 11 Comparison between measured and calculated cross-beam profiles dose distributions at two different depths (1·5 and 5·0 cm) in a water phantom caused by a 6 MV beam from an Elekta Precise SL-25 accelerator (SSD=100 cm) using the MCNPX 2·6 code. Abbreviations: FWHM, full-width at half-maximum; SSD, source and the phantom surface.

Figure 12

Figure 12 Comparison between measured and calculated depth dose distribution at the central axis of a water phantom caused by a 6 MV beam from an Elekta Precise SL-25 accelerator (SSD=100 cm) for different field sizes using the MCNPX 2·6 code. Abbreviations: FWHM, full-width at half-maximum; SSD, source and the phantom surface.

Figure 13

Figure 13 Comparison between measured and calculated cross-beam profiles dose distributions at 1·5 cm depth in a water phantom caused by a 6 MV beam from an Elekta Precise SL-25 accelerator (SSD=100 cm) for different field sizes using the MCNPX 2·6 code. Abbreviation: SSD, source and the phantom surface.

Figure 14

Figure 14 Depth dose curves in a water/bone/water phantom (left) and a water/bone/water phantom (right) calculated for a 6 MV photon beam. The field size is 10×10 cm2 at 100 SSD. Abbreviations: FWHM, full-width at half-maximum; PDD, per cent dose depth; SSD, source and the phantom surface.