In situ fracture of stents implanted for relief of pulmonary arterial stenosis in patients with congenitally malformed hearts

Published online by Cambridge University Press:  01 August 2008

Doff B. McElhinney*
Affiliation:
Department of Cardiology, Children’s Hospital, and Department of Pediatrics, Harvard Medical School, Boston, Massachusetts, United States of America
Lisa Bergersen
Affiliation:
Department of Cardiology, Children’s Hospital, and Department of Pediatrics, Harvard Medical School, Boston, Massachusetts, United States of America
Audrey C. Marshall
Affiliation:
Department of Cardiology, Children’s Hospital, and Department of Pediatrics, Harvard Medical School, Boston, Massachusetts, United States of America
*
Correspondence to: Doff B. McElhinney, MD, Department of Cardiology, Children’s Hospital, 300 Longwood Avenue, Boston, MA 02115, USA. Tel: +1 617 355 9656; Fax: +1 617 739 6282; E-mail: doff.mcelhinney@cardio.chboston.org
Rights & Permissions [Opens in a new window]

Abstract

Background

One of the most common uses of stents in patients with congenitally malformed hearts is treatment of pulmonary arterial stenosis. Although there are reports of fractured pulmonary arterial stents, little is known about the risk factors for, and implications of, such fractures.

Methods

We reviewed angiograms to identify fractures in stents previously inserted to relieve stenoses in pulmonary arteries from 1990 through 2001 in patients who also underwent follow-up catheterization at least 3 years after placement of the stent. We undertook matched cohort analysis, matching a ratio of 2 fractured to 1 unfractured stent.

Results

Overall, 166 stents meeting the criterions of our study had been placed in 120 patients. We identified fractures in 35 stents (21%) in 29 patients. All fractured stents were in the central pulmonary arteries, 24 (69%) in the central part of the right pulmonary artery, and all were complete axial fractures, or complex fractures along at least 2 planes. Stent-related factors associated with increased risk of fracture identified by multivariable logistic regression included placement in close apposition to the ascending aorta (p = 0.001), and a larger expanded diameter (p = 0.002). There was obstruction across 28 of 35 fractured stents, which was severe in 11. We re-stented 21 of the fractured stents, and recurrent fracture was later diagnosed in 3 of these. A fragment of the fractured stent embolized distally in 2 patients, without clinically important effects.

Conclusions

In situ fracture of pulmonary arterial stents is relatively common, and in most cases is related to compression by the aorta. There is usually recurrent obstruction across the fractured stent, but fractured stents rarely embolize, and are not associated with other significant complications.

Information

Type
Original Article
Copyright
Copyright © Cambridge University Press 2008
Figure 0

Table 1 Demographic and diagnostic data for patients with and without any fracture.

Figure 1

Figure 1 Fractured central right pulmonary arterial stents in two patients: (a) one with tetralogy of Fallot and pulmonary atresia, and (b) one with common arterial trunk. Both stents were posterior to the ascending aorta, and were severely compressed in the anteroposterior dimension, assuming a lenticular configuration. The stent in (a) is fractured along three axial planes: superior, inferior, and posterior, as indicated by the arrows. The stent in (b) is fractured along the superior and inferior planes (arrows). The relationship between the stent and the large ascending aorta is demonstrated by the aortogram in (b).

Figure 2

Figure 2 This stent was placed in the central left pulmonary artery, in the crook of the reconstructed aortic arch, in a patient with double inlet left ventricle who had previously undergone reconstruction of the aortic arch. (a) The native pulmonary root (P) is large and directly apposed to the stented left pulmonary artery, and the arch is pulled down and anteriorly by the reconstruction, such that the normal space beneath the arch is smaller than usual. (b) In the lateral projection, four axial fracture planes, with overlapping edges, can be appreciated clearly (arrows). The stent collapsed inward around its circumference, and the fractured edges overlap in such a manner that the stent retains a relatively cylindrical geometry, in contrast to the anteroposterior collapse of the stents in Figure 1. (c) From a left anterior oblique projection, the fracture is not as readily apparent, although disarticulated struts can be seen along the superior margin of the stent (arrows).

Figure 3

Table 2 Stent- and vessel-related factors associated with risk of fracture in case-control analysis.

Figure 4

Figure 3 (a) A central left pulmonary arterial stent in this patient with transposition who underwent repair with an arterial switch operation and anterior translocation of the pulmonary arteries is fractured along two axial planes (arrows), anteroinferior and posterosuperior, and severely compressed. (b) A pulmonary angiogram with cranial angulation demonstrates severe obstruction across the fractured and compressed stent, with obvious thinning of the contrast column within the stent, indicating a narrow anteroposterior vessel diameter. LPA, left pulmonary artery; PT, pulmonary trunk; RPA, right pulmonary artery.

Figure 5

Figure 4 This central right pulmonary arterial stent in a patient with common arterial trunk and interruption of the aortic arch is fractured along superior and inferior axial planes, but is only mildly distorted in its geometry. Over the course of a cardiac cycle, two points of which are demonstrated in (a) and (b), the torsion and compression of the vessel can be appreciated by the movement of the fractured edges of the stent (arrows) relative to one another.