Commissioning a new CT simulator II: virtual simulation software

Published online by Cambridge University Press:  01 September 2007

D. Kearns*
Affiliation:
Department of Medical Physics, Northern Ireland Cancer Centre, Belfast City Hospital, Belfast, N. Ireland
M. McJury
Affiliation:
Department of Medical Physics, Northern Ireland Cancer Centre, Belfast City Hospital, Belfast, N. Ireland
*
Correspondence to: D. Kearns, Department of Medical Physics, Northern Ireland Cancer Centre, Belfast City Hospital, Lisburn Road, Belfast BT9 7AB, N. Ireland. E-mail: denise.kearns@mpa.n-i.nhs.uk
Rights & Permissions [Opens in a new window]

Abstract

This paper continues the discussion on the commissioning tests performed on a new GE Lightspeed RT wide-bore computed tomography (CT) scanner, focusing on the GE Advantage Sim software (version 6.0).

The tests performed and phantoms used to assess the virtual simulator functionality, including the 3D image display, contouring, treatment unit beam parameters, digitally reconstructed radiograph generation and image quality, isocentre generation and multi-modality image registration, are described.

The series of tests performed showed the virtual simulation software to be working within acceptance tolerances suggested in the literature and baseline data have been obtained against which future comparisons of system performance have been made. Where no tolerances were available, we have suggested suitable values.

Information

Type
Original Article
Copyright
Copyright © Cambridge University Press 2007
Figure 0

Figure 1. Schematic diagram and CT image of the Perspex and wire cube phantom.

Figure 1

Table 1. Summary of the tests carried out to assess virtual simulator functionality

Figure 2

Figure 2. Phantom used to assess DRR ray tracing algorithm. When reconstructed at 80 cm SSD, the two markers line up as shown.

Figure 3

Figure 3. Examples of the DRRs reconstructed on Advantage Sim of the Catphan® 600 CTP528 high-resolution module and the CTP515 low-contrast module.

Figure 4

Table 2. Typical registration accuracy