Removal of a circumferentially torn stuck balloon from a calcified right ventricle to pulmonary artery conduit

Published online by Cambridge University Press:  09 January 2025

Omeir Ali Aziz
Affiliation:
University of Child Health Sciences, The Children’s Hospital, Lahore, Pakistan
Mohammad Asim Khan
Affiliation:
University of Child Health Sciences, The Children’s Hospital, Lahore, Pakistan
Masood Sadiq*
Affiliation:
University of Child Health Sciences, The Children’s Hospital, Lahore, Pakistan
*
Corresponding author: Masood Sadiq; Email: drmasoodsadiq@hotmail.com
Rights & Permissions [Opens in a new window]

Abstract

We report a case of right ventricle to pulmonary artery conduit angioplasty in which a valvuloplasty balloon ruptured circumferentially and was retained within the conduit. A high-pressure balloon was used to relieve the obstruction and free the ruptured balloon. The procedure was further complicated when the distal part of the balloon broke away from the proximal part during an attempt to retrieve it back into the femoral sheath. This report highlights the inappropriate use of a soft balloon in a calcified xenograft conduit, which led to a series of complications, and discusses potential strategies for managing these complications.

Information

Type
Brief Report
Copyright
© The Author(s), 2025. Published by Cambridge University Press
Figure 0

Figure 1. RV angiogram (full lateral) showed calcified RV–PA conduit with severe stenosis, RVH with good functions.

Figure 1

Figure 2. RV–PA conduit angiogram (Full lateral) showed partial inflation, partial relief of waist and circumferential rupture of 20 mm x 4 cm OSYPKA VACS II valvuloplasty balloon and release of contrast media in RV to PA conduit.

Figure 2

Figure 3. RV–PA conduit angiogram (full lateral) showed circumferential rupture of 20 mm x 4 cm OSYPKA VACS II valvuloplasty balloon and release of contrast media into branch PAs.

Figure 3

Figure 4. RV angiogram (AP projection) showed another super-stiff wire placed in RPA through 2nd venous access (first wire was in LPA).

Figure 4

Figure 5. RV–PA conduit angiogram (full lateral) showed inflated 20 mm x 5 cm NuMED BIB stent placement balloon catheter placed across the narrowed segment with relieve of waist.

Figure 5

Figure 6. Ruptured OSYPKA VACS II valvuloplasty balloon being removed from right femoral vein.

Figure 6

Figure 7. Circumferentially ruptured OSYPKA VACS II valvuloplasty balloon with complete removal of the ruptured fragments.