New laryngoscope for endoscopic arytenoidectomies

Published online by Cambridge University Press:  15 October 2014

T Yilmaz*
Affiliation:
Department of Otolaryngology-Head and Neck Surgery, Hacettepe University Faculty of Medicine, Ankara, Turkey
N Süslü
Affiliation:
Department of Otolaryngology-Head and Neck Surgery, Hacettepe University Faculty of Medicine, Ankara, Turkey
M D Bajin
Affiliation:
Department of Otolaryngology-Head and Neck Surgery, Hacettepe University Faculty of Medicine, Ankara, Turkey
R Ö Günaydin
Affiliation:
Department of Otolaryngology-Head and Neck Surgery, Hacettepe University Faculty of Medicine, Ankara, Turkey
S Özer
Affiliation:
Department of Otolaryngology-Head and Neck Surgery, Hacettepe University Faculty of Medicine, Ankara, Turkey
G Atay
Affiliation:
Department of Otolaryngology-Head and Neck Surgery, Hacettepe University Faculty of Medicine, Ankara, Turkey
*
Address for correspondence: Dr Taner Yılmaz, Department of Otolaryngology-Head and Neck Surgery, Hacettepe University Faculty of Medicine, 06100 Sıhhiye, Ankara, Turkey Fax: +90 312 3113500 E-mail: tyilmaz@hacettepe.edu.tr
Rights & Permissions [Opens in a new window]

Abstract

Objective:

During an endoscopic arytenoidectomy, an intubation tube must be elevated anteriorly with the laryngoscope to ensure an adequate surgical field. This paper describes a new laryngoscope that has a canal along the outer wall of the body and a ridge which runs along the canal.

Method:

Ten patients underwent endoscopic total arytenoidectomy using this new laryngoscope and 10 patients underwent the same operation using a regular laryngoscope.

Results:

The duration of all operations ranged between 25 and 65 minutes, with a median duration of 42.5 minutes. The median duration with the new laryngoscope was 39 minutes, and that with the regular laryngoscope was 49 minutes; this difference was statistically significant (p < 0.05).

Conclusion:

This new laryngoscope shortened the duration of the endoscopic arytenoidectomy and facilitated the procedure by enlarging the surgical field. This new laryngoscope may be a beneficial surgical instrument for posterior endoscopic laryngeal operations.

Information

Type
Main Articles
Copyright
Copyright © JLO (1984) Limited 2014 
Figure 0

Fig. 1 Total arytenoidectomy performed using a regular laryngoscope; the field is narrowed by the intubation tube (which is protected from the laser beam by a wet cottonoid).

Figure 1

Fig. 2 Another view of total arytenoidectomy performed using a regular laryngoscope, showing the field narrowed by the intubation tube (which is protected from the laser beam by a wet cottonoid).

Figure 2

Fig. 3 External view of the new laryngoscope.

Figure 3

Fig. 4 Closer view of the canal on the new laryngoscope.

Figure 4

Fig. 5 Internal view of the new laryngoscope; the canal does not narrow the operative field too much.

Figure 5

Fig. 6 Total arytenoidectomy performed using the new laryngoscope; the intubation tube is elevated anteriorly by the laryngoscope, making the operative field much larger.

Figure 6

Fig. 7 Another view of total arytenoidectomy performed using the new laryngoscope, showing a much larger operative field (as a result of anterior elevation of the intubation tube). The intubation tube is not in the operative field and does not need to be protected from the laser beam.

Figure 7

Fig. 8 Endoscopic suturing is much easier with the new laryngoscope; this image shows the first suture to the posteroinferior mucosal advancement flap.

Figure 8

Fig. 9 Image showing the second suture to the posteroinferior mucosal advancement flap, performed using the new laryngoscope.