Hostname: page-component-7b7567c8c-tnzcf Total loading time: 0 Render date: 2026-11-14T08:34:08.791Z Has data issue: false hasContentIssue false

Endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscope reprocessing: Variables impacting contamination risk

Published online by Cambridge University Press:  16 January 2023

Ashley M. Ayres*
Affiliation:
Department of Infection Prevention and Control, UPMC Presbyterian–Shadyside, Pittsburgh, Pennsylvania
Julia Wozniak
Affiliation:
Department of Infection Prevention and Control, UPMC Presbyterian–Shadyside, Pittsburgh, Pennsylvania
Jose O’Neil
Affiliation:
Department of Gastroenterology, Hepatology, and Nutrition, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania
Kimberly Stewart
Affiliation:
Department of Gastroenterology, Hepatology, and Nutrition, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania
John St. Leger
Affiliation:
Department of Gastroenterology, Hepatology, and Nutrition, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania
A. William Pasculle
Affiliation:
Division of Microbiology, Department of Pathology, University of Pittsburgh School of Medicine, Pittsburgh, Pennsylvania
Casey Lewis
Affiliation:
Department of Infection Prevention and Control, UPMC Presbyterian–Shadyside, Pittsburgh, Pennsylvania
Kevin McGrath
Affiliation:
Department of Gastroenterology, Hepatology, and Nutrition, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania
Adam Slivka
Affiliation:
Department of Gastroenterology, Hepatology, and Nutrition, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania
Graham M Snyder
Affiliation:
Department of Infection Prevention and Control, UPMC Presbyterian–Shadyside, Pittsburgh, PennsylvaniaDivision of Infectious Diseases, Department of Medicine, University of Pittsburgh School of Medicine, Pittsburgh, Pennsylvania
*
Author for correspondence: Ashley Ayres, E-mail: ayresam@upmc.edu
Rights & Permissions [Opens in a new window]

Abstract

Objective:

To evaluate variables that affect risk of contamination for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscopes.

Design:

Observational, quality improvement study.

Setting:

University medical center with a gastrointestinal endoscopy service performing ∼1,000 endoscopic retrograde cholangiopancreatography and ∼1,000 endoscopic ultrasound endoscope procedures annually.

Methods:

Duodenoscope and linear echoendoscope sampling (from the elevator mechanism and instrument channel) was performed from June 2020 through September 2021. Operational changes during this period included standard reprocessing with high-level disinfection with ethylene oxide gas sterilization (HLD–ETO) was switched to double high-level disinfection (dHLD) (June 16, 2020–July 15, 2020), and duodenoscopes changed to disposable tip model (March 2021). The frequency of contamination for the co-primary outcomes were characterized by calculated risk ratios.

Results:

The overall pathogenic contamination rate was 4.72% (6 of 127). Compared to duodenoscopes, linear echoendoscopes had a contamination risk ratio of 3.64 (95% confidence interval [CI], 0.69–19.1). Reprocessing using HLD-ETO was associated with a contamination risk ratio of 0.29 (95% CI, 0.06–1.54). Linear echoendoscopes undergoing dHLD had the highest risk of contamination (2 of 18, 11.1%), and duodenoscopes undergoing HLD-ETO and the lowest risk of contamination (0 of 53, 0%). Duodenoscopes with a disposable tip had a 0% contamination rate (0 of 27).

Conclusions:

We did not detect a significant reduction in endoscope contamination using HLD-ETO versus dHLD reprocessing. Linear echoendoscopes have a risk of contamination similar to that of duodenoscopes. Disposable tips may reduce the risk of duodenoscope contamination.

Information

Type
Original Article
Creative Commons
Creative Common License - CCCreative Common License - BYCreative Common License - NCCreative Common License - SA
This is an Open Access article, distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike licence (http://creativecommons.org/licenses/by-nc-sa/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the same Creative Commons licence is used to distribute the re-used or adapted article and the original article is properly cited. The written permission of Cambridge University Press must be obtained prior to any commercial use
Copyright
© The Author(s), 2023. Published by Cambridge University Press on behalf of The Society for Healthcare Epidemiology of America
Figure 0

Fig. 1.Analysis periods for duodenoscope contamination, including by method of reprocessing, change to duodenoscope with disposable tip, and change from clinical to commercial laboratory. Note. dHLD, double high-level disinfection; HLD-ETO, high-level disinfection/ethylene oxide gas sterilization.

Figure 1

Table 1.Frequency of Contamination with ≥1 Pathogenic Bacteria on Elevator Mechanism or Instrument Channel of Cultured Duodenoscopes, by Duodenoscope Type and Reprocessing Method

Supplementary material: File

Ayres et al. supplementary material

Ayres et al. supplementary material

Download Ayres et al. supplementary material(File)
File 52.9 KB