A self-made tube cracker coupled to an EA-IRMS-AGE3 system

Published online by Cambridge University Press:  20 September 2024

Lior Regev
Affiliation:
D-REAMS Radiocarbon Laboratory, Scientific Archaeology Unit, Weizmann Institute of Science, Rehovot 7610001, Israel
Eugenia Mintz
Affiliation:
D-REAMS Radiocarbon Laboratory, Scientific Archaeology Unit, Weizmann Institute of Science, Rehovot 7610001, Israel
Lilia Goffer
Affiliation:
Department of Physics Core Facilities, Faculty of Physics, Weizmann Institute of Science, Rehovot 7610001, Israel
Elisabetta Boaretto*
Affiliation:
D-REAMS Radiocarbon Laboratory, Scientific Archaeology Unit, Weizmann Institute of Science, Rehovot 7610001, Israel
*
Corresponding author: Elisabetta Boaretto; Email: elisabetta.boaretto@weizmann.ac.il
Rights & Permissions [Opens in a new window]

Abstract

The automatic graphitization system (AGE3) by IonPlus is very popular among radiocarbon dating laboratories. Usually, solid samples are burnt in an elemental analyzer (EA), and the gaseous CO2 is transferred for graphitization. Our system is coupled also with an isotope ratio mass spectrometer (IRMS), which measures the δ13C and δ15N of that gas. Some less routine pretreatment protocols require the production of gaseous samples and prevent the possibility of using the EA-AGE3 system, as the EA is used for solid samples only. In order to use that system, including the measurements of stable isotopes, we developed a glass tube cracker that connects to the EA. The device is routinely used in our laboratory and is mainly built from Swaglok catalog parts. We show that the background (blank) levels of a marble standard are indistinguishable between using the cracker and burning solid marble using the EA. We further demonstrate that the δ13C values are consistent and that the extraction efficiency when using the device is above 93%. Full descriptions, drawings, and working protocol are supplied.

Information

Type
Technical Note
Copyright
© The Author(s), 2024. Published by Cambridge University Press on behalf of University of Arizona
Figure 0

Figure 1. The cracker device. (A) The cracker device as installed and connected to the EA. Note that the additional He cylinder for washing is not connected. (B) A drawing of the cracker device. (C) A close-up image of the connection of the device to the EA columns.

Figure 1

Table 1. Part list of the cracker device. The part numbers correlate with Figure 2B

Figure 2

Figure 2. Operation modes and part numbers. (A) The three operation modes of the device are “EA Only” (blue line), “Washing” (red line), and “Collection” (yellow line). Note the ABC tags, marking the three positions of each valve. The black, blue, green, and orange tags relate to valves 1-4, accordingly. The tags are used to follow the operation protocol (online supplementary material). (B) An assembly drawing of the cracker device, including the item numbers, correlated to Table 1.

Figure 3

Figure 3. pMC levels of marbles extracted through the cracker or by the EA.

Figure 4

Table 2. Measured values of all samples

Figure 5

Figure 4. δ13C values of marble samples extracted by the cracker or EA and measured using the coupled IRMS instrument. The δ13C uncertainty is 0.2‰.

Supplementary material: File

Regev et al. supplementary material

Regev et al. supplementary material
Download Regev et al. supplementary material(File)
File 3 MB