Comparison of neutron contamination in small photon fields of secondary collimator jaws and circular cones

Published online by Cambridge University Press:  29 September 2023

Nooshin Banaee*
Affiliation:
Medical Radiation Research Center, Central Tehran Branch, Islamic Azad University, Tehran, Iran
Kiarash Goodarzi
Affiliation:
Medical Radiation Research Center, Central Tehran Branch, Islamic Azad University, Tehran, Iran
Elham Hosseinzadeh
Affiliation:
Department of Medical Physics, Faculty of Medicine, Jundishapur University of Medical Sciences, Ahvaz, Iran
*
Corresponding author: Dr Nooshin Banaee; Email: nooshin_banaee@yahoo.com
Rights & Permissions [Opens in a new window]

Abstract

Introduction:

Advanced treatment modalities involve applying small fields which might be shaped by collimators or circular cones. In these techniques, high-energy photons produce unwanted neutrons. Therefore, it is necessary to know neutron parameters in these techniques.

Materials and methods:

Different parts of Varian linac were simulated by MCNPX, and different neutron parameters were calculated. The results were then compared to photoneutron production in the same nominal fields created by circular cones.

Results:

Maximum neutron fluence for 1 × 1, 2 × 2, 3 × 3 cm2 field sizes was 165, 40.4, 19.78 (cm–2.Gy-1 × 106), respectively. The maximum values of neutron equivalent doses were 17.1, 4.65, 2.44 (mSv/Gy of photon dose) for 1 × 1, 2 × 2, 3 × 3 cm2 field size, respectively, and maximum neutron absorbed doses reached 903, 253, 131 (µGy/Gy photon dose) for 1 × 1, 2 × 2, 3 × 3 cm2 field sizes, respectively.

Conclusion:

Comparing the results with those in the presence of circular cones showed that circular cones reduce photoneutron production for the same nominal field sizes.

Information

Type
Original Article
Copyright
© The Author(s), 2023. Published by Cambridge University Press
Figure 0

Figure 1. Circular cone: a. Schematic view of a circular cone, b. circular cone attachment to the gantry.

Figure 1

Figure 2. Simulated 18 MV Varian Clinac 2100 C/D linac

Figure 2

Figure 3. Comparison of Semiflex ion chamber measurements and MC calculations for 10X10 cm2. a: PDD curve, b: beam profile

Figure 3

Table 1. Simulation validation parameters: ion chamber measurements versus MC simulation

Figure 4

Table 2. Photon absorbed dose at (dmax) for small field sizes shaped by secondary collimator jaws and circular cones7

Figure 5

Figure 4. Number of neutrons per Gy of photon dose for different field sizes along the central axis

Figure 6

Table 3. Number of neutrons per Gy of photon dose for different field sizes7

Figure 7

Figure 5. Comparison of neutron equivalent doses for various field sizes along the central axis

Figure 8

Table 4. Neutron equivalent dose for different field sizes7

Figure 9

Figure 6. Neutron absorbed dose along the central axis for different field sizes

Figure 10

Table 5. Comparison of neutron absorbed doses for different field sizes7