Field test of a new snow-particle counter (SPC) system

Published online by Cambridge University Press:  20 January 2017

Takeshi Sato
Affiliation:
Shinjo Branch of Snow and Ice Studies, National Research Institute for Earth Science and Disaster Prevention, Shinjo, Yamagata 996, Japan
Tadashi Kimura
Affiliation:
Shinjo Branch of Snow and Ice Studies, National Research Institute for Earth Science and Disaster Prevention, Shinjo, Yamagata 996, Japan
Taminoe Ishimaru
Affiliation:
Niigata Electric Company Ltd., Sawada, Nagaoka, Niigata 940–11, Japan
Toshisuke Maruyama
Affiliation:
Niigata Electric Company Ltd., Sawada, Nagaoka, Niigata 940–11, Japan
Rights & Permissions [Opens in a new window]

Abstract

The optical system of the snow-particle counter (SPC), which was developed by Schmidt in 1977, has been improved. A laser diode is used as a light source, achieving uniform sensitivity to a blowing snow particle regardless of the location of particle trajectory within a sampling volume. The light entering a slit, which may be affected by a blowing snow particle, is perfectly detected by use of a piano-cylindrical lens and a dual-type photodiode. A signal processor has been developed to get output voltage proportional to the mass flux of blowing snow.

From the estimates based on blowing snow characteristics and wind speed profile, the new SPC system can accurately detect all the particles of effective sizes at least at a height above 0.1 m when the wind speed at a height of 1 m is less than 15 m s−1.

Considering the Fraunhofer diffraction by both the wire and the particle, the relation between a particle diameter and sensor output of the new SPC system is derived from the calibration with spinning wires.

Mass flux obtained with the new SPC system was found to be close to that with a snow trap. The system was operated continuously for at least nine days using two 35 A h lead batteries.

Information

Type
Research Article
Copyright
Copyright © International Glaciological Society 1993
Figure 0

Fig.1 . Appearance of new SPC sensor. Light source in a tube is attached to the enclosure on the righthand side; and light receiver with a thin hood is attached to the enclosure on the lefthand side.

Figure 1

Fig.2 . Schematic of the light receiver.

Figure 2

Fig.3 . Block diagram of the signal processor.

Figure 3

Table 1 Summary of the particle-size distribution

Figure 4

Fig.4 . Calibration results with wires (circles) and dots (triangles). The (equivalent) particle diameter for the open symbol is calculated geometrically, while that for the solid symbol is obtained considering diffraction.

Figure 5

Fig.5 . An example of actual calibration curve (solid line). Open circles are wire-calibration data, which are plotted against the actual equivalent particle diameter defined in the text. The broken line is used in the mass flux calculation.

Figure 6

Fig.6 . Comparison of mass flux values. The bag-trap data are corrected with the collection efficiency, 0.75.

Figure 7

Fig.7 . Relation between visibility at a height of 15 m and the mass flux ratio.