Word of caution: silent late device embolisation after perimembranous ventricular septal defect closure in a 6-Kg infant

Published online by Cambridge University Press:  03 January 2024

Raymond N. Haddad*
Affiliation:
Centre de Référence Malformations Cardiaques Congénitales Complexes - M3C, Hôpital Universitaire Necker-Enfants malades, Assistance Publique – Hôpitaux de Paris (AP-HP), Paris, France
Zakhia Saliba
Affiliation:
Department of Pediatric Cardiology, Hotel Dieu de France University Medical Center, Saint Joseph University, Alfred Naccache Boulevard, Achrafieh, Beirut, Lebanon
*
Corresponding author: R. N. Haddad; Email: raymondhaddad@live.com
Rights & Permissions [Opens in a new window]

Abstract

We report on a 6-month-old infant (6 Kg/ 64 cm) with perimembranous ventricular septal defect (absent sub-aortic rim, 10 mm left ventricular entry, and 4 and 6 mm right ventricular exists) and successful retrograde closure using an 8x6 mm KONAR-MF™ VSD occluder (Lifetech, China). Immediate and 48 hours post-procedure ultrasounds showed an accurately positioned device and two jets of mild-to-moderate residual shunts. At the 2-week follow-up, the device did not change position and the shunt was stable and intra-prosthetic. The scheduled 3-month follow-up was skipped for familial reasons. The patient came back without alarming symptoms for the regular 6-month follow-up, and the device was found embolised to the left pulmonary artery. The device was retrieved surgically, and the defect was patch-closed with excellent outcomes. There was a pseudoaneurysm involving the tricuspid valve chordae and the device was endothelialized partially on one edge suggesting that embolization occurred somewhere between 3 months and 6 months post-operative. Defects with compromised anatomies should be closed surgically to avoid suboptimal results, especially in small infants.

Information

Type
Brief Report
Copyright
© The Author(s), 2024. Published by Cambridge University Press
Figure 0

Figure 1. Multiple views of baseline echocardiogram (a–d).

Figure 1

Figure 2. Discharge ultrasound showing the device sitting perfectly within the defect with a moderate residual shunt, mainly intra-prosthetic (a–d).

Figure 2

Figure 3. Two views of ultrasound follow-up at 2 weeks post-operative: device in place (a, b) with two jets of intra-prosthetic residual shunts (c, d).

Figure 3

Figure 4. Surgical removal of the embolised device from the left pulmonary artery 6 months post-operative (a, b). Note the device covered partially by neo-endothelialization (white arrow) (b).