A new hemostatic device utilizing a novel transmission structure for delivery of adrenaline and microwave energy at 5.8 GHz

Published online by Cambridge University Press:  01 June 2017

Shaun C. Preston*
Affiliation:
Medical Microwave Systems Research Group, School of Electronic Engineering, Bangor, UK Creo Medical Ltd., Chepstow, UK
Malcolm White
Affiliation:
Creo Medical Ltd., Chepstow, UK
Brian Saunders
Affiliation:
St Marks Hospital, Harrow, UK
Zacharias Tsiamoulos
Affiliation:
St Marks Hospital, Harrow, UK
Christoper P. Hancock
Affiliation:
Medical Microwave Systems Research Group, School of Electronic Engineering, Bangor, UK Creo Medical Ltd., Chepstow, UK
*
Corresponding author: S.C. Preston Email: s.preston@bangor.ac.uk
Rights & Permissions [Opens in a new window]

Abstract

A novel transmission line structure has been developed to facilitate the delivery of both adrenaline and microwave energy to achieve hemostasis. A proximal end impedance transformer and radiative tip have been designed and manufactured to provide good match between the novel hollow transmission line and the microwave source and tissue, respectively. Further consideration of the challenges and problems encountered along with evidence of successful microwave energy delivery at 5.8 GHz into porcine liver model providing a controlled and focused coagulation zone of approximately 5 mm.

Information

Type
Research Papers
Copyright
Copyright © Cambridge University Press and the European Microwave Association 2017 
Figure 0

Fig. 1. Graph showing component attenuation with variation of dielectric thickness.

Figure 1

Fig. 2. Initial prototype utilizing 90° microwave feed.

Figure 2

Fig. 3. Diagram showing the internal connections utilized for the 90° microwave feed prototype.

Figure 3

Fig. 4. Cross-section showing basic geometry of the new cable structure.

Figure 4

Fig. 5. Three-dimensional render of proximal end transformer.

Figure 5

Fig. 6. Final prototype quarter wave transformer.

Figure 6

Fig. 7. The constituent parts of the transformer.

Figure 7

Fig. 8. Photograph showing the final 3D-printed handpiece design.

Figure 8

Fig. 9. Rendering of initial radiative tip design.

Figure 9

Fig. 10. Power density plot for Macor radiative tip.

Figure 10

Fig. 11. Graph showing S11 return loss measurements for varying tip lengths from 2.5 to 3.5 mm.

Figure 11

Fig. 12. Photograph showing initial bench testing 1.

Figure 12

Fig. 13. Photograph showing initial bench testing 2.

Figure 13

Fig. 14. Coagulation zone in liver: generator set to 20 W.

Figure 14

Table 1. Approximate coagulation zones for varying power levels.

Figure 15

Fig. 15. In vivo testing photograph showing area of energy delivery.

Figure 16

Fig. 16. Photograph showing delivery of both adrenaline and energy.