Are resorbable implants about to become a reality?

Published online by Cambridge University Press:  22 March 2006

Matthias Peuster
Affiliation:
Clinic for Congenital Heart Defects and Cardiovascular Implant Research Unit, Heart and Diabetes Center, Ruhr-University Bochum, Bad Oeynhausen, Germany
Phillip Beerbaum
Affiliation:
Clinic for Congenital Heart Defects and Cardiovascular Implant Research Unit, Heart and Diabetes Center, Ruhr-University Bochum, Bad Oeynhausen, Germany
Friedrich-Wilhelm Bach
Affiliation:
Institute for Materials Science, University of Hannover, Garbsen, Germany
Hansjoerg Hauser
Affiliation:
German Research Center for Biotechnology, Braunschweig, Germany
Rights & Permissions [Opens in a new window]

Abstract

Information

Type
Review
Copyright
© 2006 Cambridge University Press
Figure 0

A corrodible iron stent crimped on a coronary angioplasty catheter (a) and after inflation to 3.5 millimetres (b). Since the stent was not stored in protective gas, there is accumulation of corrosion products surrounding the struts. After laser-cutting and electropolishing, the stent shows no corrosion and a smooth surface (c). The dilation-pressure curve shows a recoil of the stent of 2.2%, and an opening pressure of approximately 4 atmospheres (d).

Figure 1

Accumulation of iron ions after incubation of an iron sample with a diameter of 26 millimetres and a thickness of 0.91 millimetres in Ringer's solution at 37 degree Celsius. There is rapid accumulation of iron in the electrolyte.

Figure 2

Results of metabolic activity assays (WST, Roche) after incubation of human fibroblasts and human vascular smooth muscle cells with varying concentrations of ferrous iron in the growth medium. The cells show a reduced metabolic activity compared to control cells at 100-fold elevated serum concentrations.

Figure 3

The macroscopic aspect 12 months after implantation of a corrodible iron stent (left) and an American Iron and Steel Institute 316-L control stent (right) into the descending aorta of a minipig. There is no macroscopic pathology. Vascularization seems to be more pronounced on the adventitial side at the region of the iron stent.

Figure 4

Macroscopic aspects of the corrosion process after implantation into the descending aorta of minipigs after 3 (a) 6 (b) and 12 (c) months. At 3 months, the struts of the stent are clearly visible, with corrosion plaques limited to the intersections. At 6 months, there is progressive loss of structure, and accumulation of corrosion products in the wall of the stented vessel. After 12 months, there is loss of integrity. The stents are completely covered with a thin layer of neointima without excessive proliferation.

Figure 5

Expanded (a) and non-expanded (b) view of a magnesium stent (MAGIC, Biotronik). Finite element analyses show regions of stress within the stent during expansion (c) and during crimping of the stent (d). (Photographs courtesy of Biotronik).

Figure 6

Histopathology 3 months after implantation of a MAGIC stent into the coronary artery of a domestic pig. Neointimal proliferation is low.

Figure 7

The bio-Star device immediately after experimental implantation into an atrial septal defect in a sheep (top). The scaffold produced from intestinal porcine collagen is transparent. There is no evidence for formation of thrombus. The ingrowth of a conventional Starflex device is shown at bottom left, compared with an ICL device (BioStar, NMT Medical) after 3 months. (Photography courtesy of NMT medical).

Figure 8

Radiographic aspect of a “Music wire” corrosive framework of a Cardioseal/Starflex device (a). There is accumulation of corrosive wear around the centre of the device. The central wires have fractured. Ingrowth of the Music wire device was complete after 3 months, with evidence for corrosive attacks around the springs (b). (Photography courtesy of Harald Bertram, MD.)